Parameter calibration table
Sequence
No.
Default
calibration
value
Output
nominal
value
Adjust the calibration value till the output is within the
error range
00
1000
0 V DC
Zero calibration: output DC voltage
-
20 ~ 20 mV DC
01
1000
10 V DC
Offset calibration: output DC voltage 9.87 ~ 10.13 V DC
02
900
7 Vrms
Amplitude calibration: output AC voltage 6.928 ~
7.072 Vrms
03
300
0.71 Vrms
Amplitude calibration: output AC voltage 0.701 ~
0.719 Vrms
04
500
1 MHz
Frequency calibration: output frequency 1 MHz ± 20 Hz
05 ~ 24
100 ~ 150
5 Vpp
Flatness calibration: output amplitude 4.5 Vpp ~ 5.1 Vpp
5.18.3. Disable calibration
After finishing the calibration, press
【
Shift
】 【
Cal
】
key and there display 1900, press any
numeric keys then
【
#
】
key to store the calibration parameters, disable calibration and exit the
process.
During the calibration process, if wrong calibration occurred, press
【
Freq
】
key at any time to
disable calibration and exit without storing calibration parameters.
After rebooting, the generator automatically recalls and uses the calibration parameters stored
during last calibration.
5.19. Default setting
5.19.1 Continuous function of CHA and CHB
Continuous function is default after booting.
Waveform:
sinewave
Frequency:
1 kHz
Amplitude:
1 Vpp
Attenuation: Auto
Offset:
0 V DC
Duty cycle:
50%
Symmetry:
50%
Pulse width: 0.2 ms
Output mode: in-phase
Output port: open
5.19.2. Modulation function (FM, AM, PM, PWM)
Frequency deviation:
1 kHz
Modulation amplitude depth: 100%
Phase deviation:
180°
Pulse width deviation:
50%
Modulation frequency:
1 kHz
Modulation waveform:
sine
Trigger source:
internal
-25-