The allowable adjustment range is +/- 5,0°C or +/- 9,0°F.
1. Plug the thermocouple into the input connector.
2. Place the thermocouple in a known, stable temperature
environment (such as an ice dry well calibrator)
3. Allow the readings to stabilize.
4. In the setup mode (see section 4.3) for display of T1 (P 5135) or
T1 or T2 (P 5140), set the offset by pressing
„▲/HOLD“
or
„MAX/MIN/▼“
Until the primary display corresponds to the
calibration temperature.