The connection method of the serial circuit of CH1 and CH2 is as shown in the
Figure described below:
Channel Track (Dependent Mode)
In independent output mode, set the output parameters of CH1 and CH2 and switch to Channel Track mode.
When the parameters of one channel are changed, the other channel changes proportionally.
For example, in independent output mode, set CH1 voltage to 2V, current to 1A; set CH2 voltage to 4V, current to
2A. When CH1 voltage is set to 6V, CH2 voltage will be set to 12V proportionally after entering "Channel Track"
channel tracking mode. If CH1 current is set to 2A, CH2 current will be set proportionally to 4A.
Note: If the setting value is outside the nominal range, it is set to the maximum.
To set the output mode:
(1) Press the Utility key. The "Output Mode" menu is activated.
(2) Use the
▲ / ▼ keys to select the desired mode and then use the key to
select the mode.
-17-