5.4. Setting CH4 Voltage Output
1.
Connect the power supply to the power source.
2.
Press the
POWER SWITCH
[1] to turn on the power supply
3.
Press
CH2/CH4 DISPLAY SWITCH
[15] to select display CH4 voltage and current value. When
display CH4 value, the “CH4” icon will on .
4.
Press AUTO CURRENT CUT OFF
PROTECTION KEY
[27] to activate output and the
OUTPUT
INDICATOR
[25] will on.
5.
Use the
CH4 VOLTAGE KNOB
[14] to adjust CH4 voltage to give a desired output voltage.
6.
Connect the circuit to the
CH4 TERMINAL
[17, 16].
7.
When the
CH4 OVERLOAD INDICATOR
[13] is on, remove some loading on the terminal to
reduce current consumption.
Remarks:
Current output is fixed to 1A, not adjustable.
Caution:
Make sure the
INPUT VOLTAGE SELECTOR [33]
set to a correct position , otherwise; it will
damage the power supply.
Do not short the MAIN TERMINALS over 1 minute; it will damage the power supply.
-24-