5.6. Setting Series Tracking Mode
1. Press
TRACKING MODE SLECTION KEY
[23] and release
TRACKING MODE SLECTION KEY
[19] to enable series tracking mode. In series tracking mode, CH2 output voltage and current value
follows CH1 setting. The output voltage is double to the CH1 display value.
Illustration of series tracking mode
2. Turn
CH2 CURRENT KNOB [12]
clockwise to maximum current output, and then use
CH1
CURRENT KNOB
[5] adjust the desired current output value.(Reference to Section 5.2)
3. Use
CH1 VOLTAGE KNOB [7]
to adjust the desired voltage output value.
4. Connect the circuit to the
CH1 “+”TERMINAL [26]
and
CH2“-”TERMINAL [24]
to get double
voltage output.
5. For the bi-polar DC power supply with common ground, connect
CH2 “+” TERMINAL [20]
to
“GND” GROUNDING TERMINAL [22]
.
CH1 “+”TERMINAL [26]
is the positive output and
CH2 “-
”TERMINAL [18]
is the negative output.
Illustration of Bi-Polar Tracking Mode