background image

Pedigo P-2040-S Blanket Warming Cabinet Operation and Care Manual

3

O P E R AT I O N A L   P R O C E D U R E S

• L.E.D., Light Emitting Diode,

referred to is an

electronic device providing illumination.

1. TURN POWER SWITCH ON.

The heat indicator L.E.D. will illuminate and remain

lit while the unit is calling for heat.  The digital
display will indicate air temperature of the heated
compartment.  The heat indicator L.E.D. will go out
when the air temperature inside the unit reaches the
temperature set-point on the electronic thermostat.

2. VIEW OR CHANGE SET-POINT.

A. Push the SET button once.  The set-point value

(current set-point holding temperature) will be
displayed for 5 seconds.  

NOTE:

Factory default

is 90°F (32°C).

B. The set-point value can be changed by pressing

a n d   h o l d i n g   t h e   S E T   b u t t o n   f o r   a t   l e a s t   4
seconds.  The  programming mode becomes
active and the LED1 indicator will blink.  

Press

a n d   h o l d

t h e   U P o r   D OW N   a r r o w   b u t t o n   t o

change the value shown in the display.  The value
can be stored either by pressing the SET button
o r   b y   w a i t i n g   t h e   e x i t   t i m e - o u t   t o   e x p i r e
(approximately 15 seconds).  The new set-point
value will flash three times to confirm.  This value
will be retained until it is changed again.  The
minimum set-point temperature is 90°F (32°C)
while the maximum set-point temperature is
200°F (93°C).  Typical temperature settings are
around 160°F (71°C).

3. LOAD THE CABINET WITH BLANKETS.

C h e c k   t h a t   t h e   e p o x y - c o a t e d   b l a n k e t   s u p p o r t

assembly and shelf are in place.  The warming
cabinet is designed to elevate blanket temperatures
t o   a   l e v e l   w h i c h   w i l l   i n c r e a s e   p a t i e n t   c o m f o r t .
Do not fill cabinet any closer than 2" from ceiling of
warming chamber (If sensor becomes blocked,
overheating could result.)  A full load of blankets
w i l l   t a k e   s e v e r a l   h o u r s   t o   r e a c h   o p t i m u m
temperature.  Make certain the cabinet door is
securely closed after initial loading and following
each blanket removal.

DO NOT BLOCK SENSOR BY

OVERLOADING CABINET WITH

BLANKETS.  OVERHEATING

COULD OCCUR.

THERMOSTAT

and INDICATOR L.E.D. SEQUENCE

Whenever the thermostat is turned “ON,” the heat
indicator L.E.D. will indicate the power “ON/OFF”
condition of the heating cable, and consequently, the
cycling of the cabinet as it maintains the set-point
cavity temperature.  If the L.E.D. does not illuminate
after normal start-up, the main power source and
thermostat must be checked.  If the warming cabinet
does not hold the temperature as set, the calibration of
the thermostat must be checked.  If the warming
cabinet fails to heat with the L.E.D. illuminated, or
heats continuously with the L.E.D. of thermostat “OFF,”
unplug and call for service.

THERMOSTAT ACCURACY

The electronic thermostat is a precise instrument and
is designed to offer trouble-free service.  If you suspect
the temperature inside the warming cabinet does not
match the temperature indicated on the digital display,
after stabilizing, follow the instructions listed below.

1.

C h e c k   t o   m a k e   c e r t a i n   t h e   p o w e r   m e e t s   t h e
requirements of the appliance, voltage and current
rating as indicated on identification tag.

2.

Verify the temperature inside the warming cabinet
with a quality thermal indicator.

A.

W i t h   t h e   e x c e p t i o n   o f   t h e   m e t a l   s h e l v e s ,
completely empty the warming cabinet.

B.

Make certain the sensor, located inside the
warming cabinet at the left side of the ceiling,
is completely clean.

C.

Suspend the thermal indicator in the center of
the warming cabinet.

D.

Allow the temperature, set on the electronic
thermostat, to stabilize for a minimum of one
hour before comparing the digital display with
the reading on the thermal indicator.  

DO NOT OPEN THE CABINET DOOR

(

S

DURING THE

TEMPERATURE STABILIZATION PERIOD

.

If the reading on the thermal indicator does not match
the digital display within 10°F (6°C), contact Pedigo
for appropriate action.

 SET

prg

!

ERROR CODE

INDICATOR L.E.D.

HEAT

INDICATOR L.E.D.

CHAMBER

TEMPERATURE

L.E.D.

TEMPERATURE

SET

BUTTON

POWER
ON/OFF

ROCKER SWITCH

L.E.D.1

INCREASE / DECREASE

BUTTON

200

THERMOST

A

T

"I" = "ON"

   "

" = "OFF"

CONTROLLER

Summary of Contents for P-2040-S

Page 1: ...P 2040 S Blanket Warming Cabinet Manual...

Page 2: ...Heat system incorporating low watt density cable mounted around each compartment This system provides evenly distributed heat without the need of a heat circulating fan Each compartment is controlled...

Page 3: ...NOT BLOCK SENSOR BY OVERLOADING CABINET WITH BLANKETS OVERHEATING COULD OCCUR THERMOSTAT and INDICATOR L E D SEQUENCE Whenever the thermostat is turned ON the heat indicator L E D will indicate the p...

Page 4: ...oor gaskets are kept free of any debris that may accumulate Good equipment that is kept clean works better and lasts longer CLEAN THE UNIT REGUL ARLY 1 Disconnect the cabinet from the power source 2 R...

Page 5: ...heck wires for integrity Check for proper and secure connections at the thermostat and terminal block If necessary re secure the faulty connections Energize system after the above steps have been comp...

Page 6: ..._____________________ ____________________________________________________________________________ ____________________________________________________________________________ ________________________...

Page 7: ...Y 2 RL 3736 12 TRANSFORMER 2 TN 3972 13 FUSE 1 4 A SLOW BLOW 2 FU 33332 14 SENSOR SERVICE KIT 2 14785 15 TERMINAL BLOCK SENSOR 2 BK 33546 16 TERMINAL BLOCK 2 BK 33364 17 SLAB DOOR ASSEMBLY 2 15147 18...

Page 8: ...Pedigo P 2040 S Blanket Warming Cabinet Operation and Care Manual 8 P 2040 S Bonnet Service View TT 33563 w E3037TT SN 33540 BK 33546 BK 33364 NOT SHOWN TN 3972 FU 33332 E3006FU TT 33563 w E3037TT...

Page 9: ...ation and Care Manual 9 P 2040 S Cabinet Service View E2088PE GS 23794 E4015 E1006 SN 33540 SW 3962 TT 33563 w E3037TT 4460 1844 1897 CASINGS 15147 HD 24171 4265 4572 CS 2026 FRONT CASTERS SWIVEL w BR...

Page 10: ...Pedigo P 2040 S Blanket Warming Cabinet Operation and Care Manual 10...

Page 11: ...hever occurs first The parts warranty on all other parts remains in effect three 3 years from installation or thirty nine 39 months from the shipping date whichever occurs first This warranty does not...

Reviews: