background image

INT - Measuring interval

Defines the time interval for temperature and humidity measurement in the 

range of 5 - 120 minutes.

C- F - Temperature unit

Changes the temperature unit from °C to °F. All settings (e.g. alarm limits) as 

well as recorded values are displayed according the selected unit. Data on 

the microSD-card is also stored in the selected unit.

IR - Infrared Receiver

Sets MostraLog’s remote control response time in the range of 1 - 5 

seconds. If, for example, a reaction time of 5 seconds is selected, keys on 

the remote control must be pressed for at least 5 seconds to be recognized. 

Short intervals improve reaction time but considerably reduce battery life. 

If no remote control is used, select “OFF“ for maximum battery life (factory 

setting). When the IR receiver is switched off, you can operate the MostraLog 

via the device keys or via the PC software.

Note: Some light sources (e.g. energy-saving lamps) generate flicker in the 

non-visible IR range. This can make MostraLog constantly detect false IR key 

presses and thus very quickly drain the device’s battery. Permanently deacti-

vate the IR receiver if the batteries are completely discharged after only a 

few days.

LCD - Display contrast

Adjusts the contrast of the LCD display. Contrast is influenced by the 

ambient light conditions and the reading angle. Press the “

“, button, the 

display shows “OK“. 

Note: The weaker the batteries, the lower the display contrast. If the battery 

charge level falls below 30%, the display contrast is set to maximum to 

ensure readability even with weak batteries.

CAL - Calibration

A calibration set with two saturated calibration solutions (sodium chloride 

and magnesium dichloride) is available for readjusting the MostraLog. 

Detailed calibration instructions are enclosed with the calibration set. Select 

the appropriate calibration point: NA (NaCl) or MG (MgCl2) and start the 

calibration process with “

“. The display shows “OK” after successful calibra-

tion. If an error occurs, the display shows “ERR“. Possible causes of errors are, 

for example, contaminated calibration solutions or wrong media selection 

(NaCl instead of MgCl2).

DEF - Factory setting  (default)

Resets the unit to the factory settings. Note that the IR receiver is turned off 

by default. Date and time will not be changed.

VER - Firmware version

Displays the firmware version on the device.

Configuration with PC software

When the device is connected to a PC by mini-USB cable, data can be read 

out and the device can be configured with the MostraLog PC software. 

Since the device is connected to the PC as a mass storage device, you can 

also access the data using Windows Explorer. To do so, wait until “MostraLog 

Device“ and a drive letter appear. Then you can access the device like a USB 

stick and transfer the recorded data to the PC.

Note: If the device is connected to a PC via USB, the display is cleared and 

only the message “USB“ is displayed to indicate PC control.

Note: Depending on the capacity of the microSD-card, it may take several 

minutes for the drive letter to appear on the PC. Use low-capacity microSD-

cards to minimize access times.

Batteries and power supply

Note: Almost all instrument settings are retained after a battery change. Only 

the date and time must be readjusted, either by connecting the device to 

your computer (MostraLog takes the computer’s time) or manually via the 

device menu. 

MostraLog can also be powered via a USB power supply. In order to do 

this, connect MostraLog to a standard USB power supply using a mini-USB 

cable. Any batteries should be left in the device, but won’t discharge when 

supplied via USB. The display shows “USB“ instead of battery charge status. 

The USB power supply makes it possible to change batteries without 

interrupting data recording. USB-power the device before removing the 

batteries, and leave it connected while you replace the batteries. After you 

have inserted the new batteries, the USB power supply can be disconnected.

It is also possible to connect the MostraLog permanently via USB to use the 

batteries as backup batteries, e.g. during a power failure.

External temperature/humidity sensor

If an external sensor is connected, the display shows values from the external 

sensor, otherwise the values of the internal sensor are displayed. Values from 

both sensors are stored on the microSD-card in parallel, resulting in four data 

tracks (internal/external temperature, internal/external humidity).

Technical data

Temperature 

sensor

Measuring range

-18°C to +70°C (0°F to 158°F)

Accuracy

±0.5°C(5°C to 45°C); 

 ±1°F (41°F to 113°F)

Resolution

0.1°C (0.2°F)

Humidity 

sensor

Measuring range

0 to 100% RH

Accuracy

±2% RH (10 to 90% RH)

Resolution

0.1% RH

Operating range 

-18°C to +70°C (0°F to 158°F)/  

0% to 100% RH

Memory

2 GB microSD-card

Measuring interval  5 - 120 min

Start options 

Immediately

Alarm

Min-/Max limits, low battery

Alarm output 

low side switch, digital signal

Data transfer 

USB or microSD-card

Battery

2 x standard AAA, replaceable

Battery life 

Up to 2.5 years at 23°C (73°F)

Housing 

ABS, black or white

Dimensions 

58 x 47 x 21 mm (2,3 x 1,8 x 0,8 inch)

Weight 

57 g (2 oz)

External sensor 

Cable 2m or 5 m; sensor head Ø 8 mm

Reviews: