CAUTIONS AND WARNINGS 

INTENDED USE 

This Health o meter

®

 Professional platform scale is intended to be used in a professional medical environment 

by trained medical staff. This product was designed to weigh patients who are safely positioned and standing 
on the platform. If a height rod is attached to the scale, the intended use of the height rod is to measure patient 
height. Do not modify the product or use it for anything other than its intended purpose. 

 

To prevent patient/caregiver injury or damage to your scale, please follow the instructions in this 
user manual very carefully. 

 

 

Do not exceed the weight capacity specified for this scale. 

 

For accurate weighing, this scale must be placed on a flat, stable surface. 

 

For accurate weighing, verify before each use the proper operation according to the procedure 
described in this manual. 

 

Do not use in the presence of flammable or explosive materials. 

 

If the scale becomes damaged, it should not be operated until properly serviced.  

 

Do not wheel or pull the scale down stairs, doing so may damage internal parts. 

 

Ensure that the AC adapter and scale do not come in contact with liquids, excessive temperature, 
or excessive humidity. 

 

Patient/caregiver safety 

 

This scale is designed for static weighing of patients only. No scale should be used for patient 
transfer.  

 

To prevent patient injury, the patient must be attended throughout the entire weighing event.  
Caregivers should ensure that the patient is stable and provide support as needed when getting 
onto and off of the scale. 

 

Patient should wear socks or other form of lightweight foot covering during weighing event. 

 

 When operating the scale with the optional AC adapter: 

 

Operate this scale exclusively with the Heath o meter Professional  AC adapter  (ADPT30). Use 
of an unspecified adapter will void the warranty, and can pose a serious safety hazard.   

 

Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent 
plug prongs. 

 

Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated 
voltage appropriate for operation.   

 

Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.  

 

Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the 
scale.   

 

Operating this scale at voltages and frequencies other than specified can damage the equipment 
and will void the warranty. 

 

When operating the scale with batteries: 

 

If the scale’s battery icon indicates that the batteries are low, replace the batteries or switch to the 
AC adapter as soon as possible.  

 

When replacing / inserting batteries, be sure to use all new batteries. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 
93/42/EEC (EN 55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable 
protection against harmful interference in a typical medical installation. The equipment generates, uses 
and may radiate radio frequency energy and, if not installed and used in accordance with these 
instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that 
interference will not occur in a particular installation. If this equipment does cause harmful interference with 
other devices, which can be determined by turning the equipment off and on, the user is encouraged to try 
to eliminate the interference by one or more of the following measures: 

 

Reorient or relocate the receiving device. 

 

Increase the separation between the equipment. 

 

Connect the equipment into an outlet on a circuit different from that to which the other device(s) is 
connected. 

 

Consult Customer Service or a field service technician for help. 

 

In no event whatsoever shall Pelstar

®

, LLC be liable for damages or injuries whatsoever arising 

out of or connected with the assembly, use, or misuse of its products. 

Summary of Contents for Health O Meter 3105KG-AM

Page 1: ... Platform Scale User Instructions Para ver las instrucciones de uso en español visita www homscales com PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 3105KL AM Health o meter Professional 3105KG AM Pelstar LLC 2022 Equipped with Everlock ...

Page 2: ...nd Warnings 3 Specifications 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Set Up Instructions 10 Everlock Instructions 13 Operating Instructions 14 Maintenance and Cleaning 23 Calibration 24 Troubleshooting 25 Warranty 26 Note This scale has been factory calibrated and does not require calibration prior to use Please register your scale for warranty coverag...

Page 3: ...y and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source Unplug the AC adapter and caref...

Page 4: ...105KG AM 454 kg x 0 1 kg Power Requirements Adapter Model No UES06WNCP 090050SPA Part No UE190716HKSH2RM Health o meter Professional Model ADPT30 not included Input 100 240VAC 50 60Hz 500mA Output 9 0V 0 5A or 6 D batteries included Environmental Operating temperatures 50 F to 95 F 10 C to 35 C Storage temperatures 30 F to 125 F 0 C to 50 C Maximum Humidity 85 RH Scale information can be found by ...

Page 5: ... Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Note If this product was purchased as a BT model refer to the Wireless Communication Set Up Instructions included with the packaging Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or local regulations which apply to you...

Page 6: ...acking materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts List 1 Scale base 1 Handlebar 1 Pillar with display head attached 6 D cell batteries 1 User Instructions Hardware 2 3 8 24 x 2 cap screws and washers 8 5 16 24 x 5 8 cap screws Tools Required Phill...

Page 7: ...with display head Align the pillar above the base with the front of the display head facing the platform Locate the blue load cell cable protruding from the bottom of the pillar 5 Plug the load cell cable into the load cell port located on the scale base 6 Slide the pillar down into the base Caution Ensure the cable is not pinched inside the pillar ...

Page 8: ... through the other mounting hole 8 Obtain the handlebar and align the holes in the handlebar above the holes in the platform Note Ensure the handlebar is mounted in the correct orientation The rear of the handlebar has the angled post refer to photo 9 Obtain the eight 5 16 24 x 5 8 cap screws Insert a screw through the handlebar and into the platform on the top outside holes on both left and right...

Page 9: ...the power source Use a Phillips head screwdriver to remove the four screws from the battery door on the rear of the pillar 2 Gently rest the battery pack on the outside of the scale Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professional recommends using an alkaline battery rated for 130 F 54 C operation Do not use Zinc Carbon batteries 3 ...

Page 10: ...ure is not available on KG models If the scale is locked per the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button to briefly display the weight in the alternate weight measuring unit After releasing the Unit button the display will return to the locked weight measuring...

Page 11: ...the next section to set the preferred time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll through the options until AC Auto Off is highlighted in red 4 Press to change the AC Auto Off option Choose Enable for the display to turn off afte...

Page 12: ... from sleep mode will not prompt the scale to zero Disable means the scale does not attempt to find a zero when powered on rather it will use the last zero setting that was obtained from the last push of the zero button There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that if the Disable option is selected users should manually zero the ...

Page 13: ...essional Scales Customer Service at 1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll through the options until Everlock is highlighted in red 4 Press to begin the Everlock procedure To exit the procedure without engaging Everlock press 5 The screen will p...

Page 14: ... seconds prompts entry of a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View setting is enabled press and hold Unit button to briefly display the alternate weight measuring unit Note The Unit button is disabled on KG models or on KL models with Everlock...

Page 15: ... s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 4 To reweigh the patient without having them step off of the scale press 5 To see the patient s weight in another unit of measurement press and hold down If the scale is in LB mode it will display the weight to KG If the scale is in KG mode it will display the weight t...

Page 16: ...e to lock onto the weight 3 After obtaining the patient s weight and with the patient remaining on the scale press 4 The scale will prompt entry of the patient s height Use the arrow keys to enter the patient s height Note When weighing in pounds the height is displayed in inches When weighing in kilograms the height is displayed in centimeters 5 Once height is keyed in press 6 The display will sh...

Page 17: ...ll instruct the user to reset the height rod to its home position or prompt the user to press an arrow key to enter height manually Push the height rod down to its home position When at home position the display will read 3 ft 11 2 inches when weighing in pounds or 120 cm when weighing in kilograms Move to step 5 Note Resetting the height rod to the home position is only needed once after a power ...

Page 18: ...t of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as a negative amount The tare value will remain on the display until changed cleared or until the scale is powered off Note When in Tare mode the screen will not show a prompt to enter height mode 6 The weight may be ...

Page 19: ... will show the patient s weight Note When in Tare mode the screen will not show a prompt to enter height mode 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 6 To reweigh the patient without removing them from the scale...

Page 20: ...d weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note When in Tare mode the screen will not show a prompt to enter height mode Note For an alternate method after taring the parent s weight in step 3 have the parent step off of the scale With no weight on the scale the display will show the t...

Page 21: ...e to its desired location 4 Gently lower the scale s platform onto the floor to avoid damaging the load cells Note If underweight message appears press CAUTION To prevent damage do not drop the scale or let the platform strike harshly against the floor when lowering Connectivity Information To enable reliable transmission of weight height and Body Mass Index BMI data this scale is designed to conn...

Page 22: ... meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild soap and water solution 70 isopropyl alcohol solution with 1 5 hydrogen peroxide concentration After cleaning disinfecting wipe with a cloth dampened with water and then with a clean dry cloth To avoid residue build up or damp surfaces ensure the screen and scale parts are completely dry aft...

Page 23: ...ocked unit of measure Note Only weights certified and traceable to national standards should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to select Calibrate Press to advance to the next step 3 The display will prompt for entry of the password Press the foll...

Page 24: ...y will prompt for the calibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is successful the display will show Calibration Complete with a 4 digit calibration count 11 If calibration is unsuccessful the display will show an error message that the entered calib...

Page 25: ...ent is not standing still Ask the patient to stand still Scale shows a weight upon power up and cannot be cleared by pressing The scale does not have a proper zero point Power off the scale Then power on the scale and immediately press and hold Release the button when the scale beeps Scale will not register a weight 1 Load cell cable is not plugged in or damaged 2 Shipping bolts were not removed 1...

Page 26: ... flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make sure you keep your sales receipt or document showing proof of purchase Call 1 800 638 3722 or 1 708 377 0600 to receive a return authorization RA number whi...