background image

Cascade Comfort System  

 

Manufacturer:  Penner Manufacturing Inc 

101 Grant St 

/ PO Box 503 

Aurora, NE 

68818 

(402) 694-5003 

 

   

 

 

Models:   Cascade – Comfort 

 

560010-X 

Cascade Comfort bathing system Right hand entry 

560010-XL 

Cascade Comfort bathing system Left hand entry 

560010-XSB 

Cascade Comfort bathing system Right hand entry Still Bath 

560010-XLSB  Cascade Comfort bathing system Left hand entry Still Bath 

(X in the model number represents a number which denotes a Color of the Spa) 

 

Spa Electrical Ratings: 

120 Volts AC 

60 Hz  

 

1 Phase 

Aqua-Aire:   

9.25 amp 

60 Hz  

 

1 Phase 

UL Classification: 

Class I; Type B; Ordinary EQUIPMENT 

 
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with 
oxygen or nitrous oxide.

 

 
MEDICAL EQUIPMENT 
WITH RESPECT TO ELECTRIC SHOCK 
FIRE AND MECHANICAL HAZARDS ONLY 
IN ACCORDANCE WITH UL2601-1, AND  
CAN/CSA C22.2 NO. 601.1 IEC 60601 
 

Patient weight rating: 400 lbs Maximum  

Upright position to reclined position angle: approximately 30°  

Reclining actuator rating: 1 minute ON - 9 minutes OFF 

 

All temperature read-out values are accurate to +/- 2ºF (per NSF C-2). The read-out range is  
-40/199ºF. The resolution is 1º. 

An electrical circuit diagram can be found on the inside of the lid of the electrical box located inside the service 
access door on the right side of the spa cabinet.  
Service parts and technical information are available by contacting your distributor or by calling one of the 
telephone numbers located on the front or back cover of this manual. 
Environmental conditions for transport and storage for up to 15 weeks (as packaged from the manufacturer) are 
as follows: 
 

Temperature range 

 

-40º F (-40º C) to 120º F (49º C) 

Relative Humidity 

 

10% to 90% 

Atmospheric Pressure   

7.5 PSI to 15.5 PSI 

Basic Components 

The Cascade Comfort System is composed of the following components: 
Cascade Spa with Cascade Door and Control Panel Cabinet. 

 

 

 

The remainder of this section describes the components and their relationship to each other. 

 

 

61LN 

 

Summary of Contents for Cascade Comfort 560010-X

Page 1: ...e Operation Daily Maintenance Instructions PENNER PATIENT CARE INC Box 523 101 Grant St Aurora NE 68818 560750 Revision A 12 11 15 1 866 PENNERS 1 866 736 6377 1 800 732 0717 E mail Address pennersale...

Page 2: ...COMFORT SPA CONTROLS 5 CASCADE DOOR 6 LOCKING CASCADE DOOR 6 UNLOCKING CASCADE DOOR 6 SYSTEM PREPARATION BEFORE BATH 7 USING CASCADE COMFORT BATHING SYSTEM 8 BATHING PROCEDURE 8 TRANSFERRING OUT OF C...

Page 3: ...tion information Failure to obey a caution warning may result in damage to the equipment and may void the warranty Left or Right When the terms left or right are used with reference to the spa this me...

Page 4: ...tient weight rating 400 lbs Maximum Upright position to reclined position angle approximately 30 Reclining actuator rating 1 minute ON 9 minutes OFF All temperature read out values are accurate to 2 F...

Page 5: ...ECT JETS 3 HAND SPRAYER ON AUTO OFF 4 FILL SHOWER TEMPERATURE DISPLAY 5 FILL SHOWER TEMPERATURE CONTROL MIXING VALVE 6 SPA TEMPERATURE DISPLAY 7 AQUA AIRE ON OFF 8 TUB FILL ON OFF 9 HAND SHOWER 10 SPO...

Page 6: ...cade Door Figure 2 1 With door handle in the upright position close the door tight against the spa 2 When the door is against the spa push the door handle to the left or down so the door handle is in...

Page 7: ...uch as a washcloth towels and liquids such as shampoo and body wash You are now ready to bring the Resident or Patient to the Bathing Area with your Patient Transfer System or Wheel Chair 2 At the bat...

Page 8: ...st the water temperature by turning the Temperature Control knob to normal bathing temperature 95 to105 degrees Fahrenheit 35 to 41 degrees Celsius 2 When water reaches the overflow push Tub Fill butt...

Page 9: ...he sprayer into the spa not at the resident Press the trigger on the hand sprayer and adjust the mixing valve if needed to the normal bathing range of 95 to 105 degrees Fahrenheit 35 to 41 degrees Cel...

Page 10: ...rain 5 Press and hold the Disinfect Button located on the control panel As the button is held down the properly mixed cleaning solution is running through the air injection system and out all of the a...

Page 11: ...isibly check that the spa was effectively cleaned during the disinfecting procedure If not repeat the procedure 14 If there is a delay of one or more hours before the next bath we recommend using a to...

Page 12: ...nfectant concentrate from damaging their eyes or skin If disinfectant gets on the skin or in the eyes rinse thoroughly with plenty of water Seek medical advice if irritation occurs 2 Perform the Daily...

Page 13: ...door WARNING If during the safety checks you find parts are missing are excessively worn do not function properly or do not meet the recommended safe operating levels do not operate the equipment unt...