BARCON LHC/PPC
Operation of devices with HART
®
communication
Subject to reasonable modifications due to technical advances.
Copyright Fuchs, Printed in Germany
Fuchs Group • Tel.: Germany (06 21) 7 76-0 • USA (330) 4 25 35 55 • Singapore 7 79 90 91 • Internet http://www.pepperl-fuchs.com
D
a
te
of
I
s
s
u
e
1
5.
1
2
.2000
44
9.2.3
Register card - General parameters
Polling address
Address of the BARCON device in the "short integer" format.
Measuring piont
• Tag:
Name of measuring point
• Measuring point description:
Information about measuring point
• Parameterization dated on:
Date of last parameterization
Units - measured values
Units of the measured values:
• mbar, bar PSI, at, kg/cm², mA, %, mm, m, inch, feet, Pa, kPa, Mpa, mmWS, mmHG
Volume reference units
• l, kg, t, m³, gal, lb
Reference value
The reference value of the volume-based units is only active when the volume units
have been activated.
• 100% = 0.0 (value range 0 ... 3000.0).
When displaying or calibrating values in units of height (e. g. mm, m,
feet, inches), the corresponding value of the density of the medium
must be entered to enable conversion to the correct filling level (see
also section 9.2.6).