Operation
7
Operation
7.1
Read data
The Handheld cannot read the data stored on a data carrier until you select the correct setting
for the data carrier type, known as the
Tag type
.
The read data is saved in the Handheld or sent to the computer depending on the setting
Settings > Send/Log
(see chapter 6.3).
If you wish to add a time stamp to the read data, activate the menu entry
ON
in the
Settings >
Time stamp
menu.
Address
In the data input field of the
Address
menu entry, you have the option of changing the start
address.
Start addresses specify the location of the individual data blocks on data carriers. You can read
out specific data blocks and transfer data by changing the start address.
Defining a start address
Define the start address as follows:
1.
Select
Address
.
2.
Enter the start address using the input keys.
3.
Press the left softkey (
) to confirm your entry.
Setting the number of blocks
The number of blocks determines how many data blocks the Handheld reads from the data
carrier and how many data blocks can be written to the data carrier. Each block corresponds to
4 bytes.
1.
Select
Read/Write > Data read > No. blocks
.
2.
Refer to the table to calculate how many blocks you require to enter the relevant data.
3.
Press the CLEAR input key (
) to delete the quantity already entered.
4.
Enter the number of blocks using the input keys.
5.
Press the left softkey (
) to conform your entry.
Read/Write
/
Storage
Settings
Applications
About
Options
Address
No. blocks
write
ESC
Data
Data read
Data write
Data copy
Read fixcode
Write fixcode
Read register