4.1 POWERUP

Press the     key to turn the thermometer On or OFF.

4.2 CONNECTING THE THERMOCOUPLES

For measurement, plug the thermocouple into the input connectors.

4.3 SELECTING THE TEMPERATURE SCALE

When the meter was first powered on, the default scale setting is set 

at Celsius (°C) scale. The user may change it to Fahrenheit (°F) by 

pressing “ °C/°F ” button and vice versa to Celsius.

4.4 DATAHOLD OPERATION

The user may hold the present reading and keep it on the display by 

pressing the “HOLD” button. When the held data is no longer needed, 

one may release the data-hold operation by pressing “HOLD” button 

again. When the meter is under Data Hold operation, the “

REL”             

,              and “ °C/°F ” button are disabled.

4.5 T1,T2,T1T2 DISPLAY CONTROL

One may select T1,T2 or T1-T2 to show on the main display by 

pressing         button. When T1 or T2 is select to show on the main 

display, the other temperature will be shown on the second display. 

When one select T1-T2 to show on the main display, T1 and T2 will be 

shown on the second display alternately.

4.6 RELATIVE OPERATION FOR MAIN DISPLAY

When one press the “

REL” button, the meter will memorize  the  

present    reading  and  the    difference  between the new reading and 

the memorized data will be shown on the display. Press the “

REL” 

button again to exit the Relative operation.

 

 

4. OPERATION INSTRUCTIONS

P.7

Summary of Contents for TC0301

Page 1: ...Datalogger Thermometer TC0301 Instruction Manual...

Page 2: ...1 Power Up 4 2 Connecting the Thermocouples 4 3 Selecting the Temperature Scale 4 4 Data Hold Operation 4 5 T1 T2 T1 T2 Display Control 4 6 Relative Operation 4 7 MAX MIN AVG Operation 4 8 Auto Power...

Page 3: ...ype thermocouples NUMERICAL DISPLAY 4 digital liquid crystal display MEASUREMENT RANGE 200 C 1370 C 328 F 2498 F RESOLUTION 200 C 200 C 0 1 C 800 C 1370 C 1 C 200 F 200 F 0 1 F else 1 F MAXIMUM VOLTAG...

Page 4: ...ement the accuracy is 0 5 T1 T2 reading 2 C or 0 5 T1 T2 reading 2 F TEMPERATURE COEFFICIENT For ambient temperatures from 0 C 18 C and 28 C 50 C for each C ambient below 18 C or above 28 C add the fo...

Page 5: ...Carrying Case Instruction Menu OPTION Soft Ware Package Program RS232 Connection Cable AC Adapter POWER REQUIREMENT 9 Volt Battery NEDA 1604 or JIS 006P or IEC6F22 BATTERY LIFE Approx 100hrs with alka...

Page 6: ...e display data is being hold The Maximum value is now being displayed The Minimum value is now being displayed The Average value is now being displayed The reading is now under Relative Mode The Batte...

Page 7: ...ensor connector LCD display ON OFF button HOLD button T1 T2 T1 T2 control button Relative readout button MAX MIN Average control button C F control button Offset calibration screw 11 Digital output co...

Page 8: ...a hold operation by pressing HOLD button again When the meter is under Data Hold operation the REL and C F button are disabled 4 5 T1 T2 T1 T2 DISPLAY CONTROL One may select T1 T2 or T1 T2 to show on...

Page 9: ...her This means that all these data is updated in the memory and the reading is the present temperature One may press to circulate the display mode among these options When the meter is under operation...

Page 10: ...orming offset Calibration with thermal stabled ice water through VR1 will give a very good calibration result 4 11 DIGITAL OUTPUT The Digital Output is a 9600bps N 81 serial interface The RX is a 5V n...

Page 11: ...RS232 command K ASC 4BH D ASC 44H B ASC 42H S ASH 53H H ASC 48H T ASC 54H M ASC 4DH N ASC 4EH R ASC 52H C ASC 43H A ASC 41H Function Ask for model No Ask for main display Range Data Unit Ask for secon...

Page 12: ...n mode HOLD MAX REL if the mode is not entered the related characters will be left as space For example when the meter is under MAX display the meter will return MAX COMMAND T Equivalent to one pushin...

Page 13: ...C 0 F 3nd BYTE bit0 1 main window value is OL 0 not OL bit1 1 main window value is minus 0 main window value is plus bit2 1 4th byte and 5th byte represent 0 4th byte and 5th byte represent bit3 1 su...

Page 14: ...window is T2 11 Main window is T2 sub window is T1 4th BYTE first two BCD code of main window value 5th BYTE last two BCD code of main window value 6th BYTE first two BCD code of sub window value 7th...

Page 15: ...r products and services please send us an email cs perfectprime com For B2B or project based application please send an email sales perfectprime com FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISI...

Page 16: ...HORMES LIMITED cs perfectprime com G F UNIT 3 61 GLENTHORNE ROAD LONDON W6 0LJ UNITED KINGDOM 44 203 7695377 Retailer Email Address Telephone...