16
Application
Safety instructions
Important: prior to starting initial operations,
be sure to read the General Information,
Point 1, at the beginning of these Instruc-
tions for Use. Improper use can result in
damage to the equipment.
Check package content
– Measuring instrument
– Lancehead-shaped probe
– Plug-in power supply (12V/2A)
– Charging adapter
– Protective cover
– Instructions for Use
Charging the battery
Charge the fitted battery before the first
use. To do this, plug the supplied charging
adapter into the 7-pin socket on the hand-
held measuring instrument. Then connect
the plug-in power adapter to the charging
adapter. If the hand-held measuring in-
strument is already switched on or the
battery is deep-discharged, the charging
process will start immediately. Other-
wise, turn on the hand-held measuring in-
strument by pressing the “Measure” but-
ton for approx. 1 second. An active charge
is indicated on the display by an animated
battery icon.
SONO WZ
Translation of the Original Instructions for Use
The integrated charging electronics
charge the battery until it is fully charged.
The complete procedure takes about 2
hours. As soon as charging is finished, all
4 “battery bars” appear permanently in
the display and trickle charging then be-
gins.
Only charge the battery at room tempera-
tures (approx. 10 °C to approx. 30 °C).
If the temperatures are too low, the end-
of-charging shutoff may not operate safe-
ly and the battery could be overcharged. If
the ambient temperatures are too high,
the hand-held measuring instrument may
be damaged by the heat generated during
charging
Connecting the probe
The hand-held measuring instrument is
operated with the lancehead-shaped
probe:
The lancehead-shaped probe is connect-
ed to the hand-held measuring instrument
via the 7-pin connector. The union nut is
then tightened.
Initial Operations