Analytes
The stand-alone mode has several predefined assays.
The other assays can be added and defined with the help of your local PerkinElmer representative.
Physical dimensions
Height: 300 mm (~11.81 in)
Width: 876 mm (~34.49 in)
Depth: 740 mm (~29.13 in)
Weight: 45 kg (~99.21 lbs)
These dimensions are excluding the PC and screen. The touch screen can be placed on the top of the instrument.
Power requirements
Power consumption: 70 VA
Mains Voltage: 100-240 V, 50-60 Hz
Environmental conditions
The Panthera-Puncher 9 is a benchtop instrument designed to be used in a normal laboratory environment. It should
be laid on a flat surface, such as a table or a workbench, and if possible, keep it away from direct sunlight as extraneous
light can interfere with barcode reading. Under such circumstances, the lights of the Panthera-Puncher 9 might need to
be adjusted.
Operating temperature: 15-30 ºC
Relative humidity: 20-80 % for punches of 3.2 mm or greater, 40 - 80 % (not condensed) for 1.5 mm punches
When operating at less than 20% relative humidity, static electricity increases the probability of disks jumping out of the
wells.
EMC
Electromagnetic compatibility according to IEC/EN 61326-2-6. Emissions according to Class B limit values.
Packaging, storage, and transportation requirements
Temperature
Storage: -20 to 50º C
Transportation: -20 to 50º C
Relative humidity (not condensed)
Storage: 5% to 90%
Transportation: 5% to 90%
Conformance to EU directives
The CE mark conforms to the following EU directives:
• 2014/30/EU (Electromagnetic Compatibility Directive)
• 2014/35/EU (Low Voltage Directive)
Panthera-Puncher
™
9 Instrument Manual
Specifications