Installation
102
Filling the PolyScience Chiller
Use the following procedure to fill the PolyScience
®
WhisperCool
TM
Chiller, and
refer to the following figure.
1. Turn the filler cap counter clockwise and lift the cap off.
2. Fill the reservoir with the coolant fluid (Part No. N0776099). The chiller
reservoir has a capacity of 4.2 liters (1.1 US gallons).
3. Keep some coolant fluid beside the unit and proceed to
Starting Up the
PolyScience Chiller
. The reservoir tank may need to be topped off once the unit
has been started.