Installation
104
1. Turn the On/Off switch to the On position. The refrigeration system and the
recirculation pump will start.
2. With the chiller running, top off the reservoir tank and screw in the reservoir
plug.
3. Install the reservoir access panel and secure it with its screws.
Connecting the System Components
The system component connections are as follows (see the following figure):
•
IEEE cable from the computer to the instrument IEEE connector on the
spectrometer side of the instrument. IEEE cable from an optional
autosampler to the instrument IEEE connector.
Note: Extending IEEE cables by making tandem connections will compromise instrument
communication.
•
The Optima 8000 is equipped with an IEC 309 250 V 16/20A 2 pole plus
protective earth plug (PerkinElmer Part No. 09997530) that inserts into an
equivalent IEC 309 series receptacle (see Section 2, Electrical
Requirements).
•
Connect the computer and printer to electrical power as described in the
guides provided with these items.
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power
supply before you perform the procedures described below.
Note
Make sure that the computer and printer share a common ground with the
instrument
.
!
Warning
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......