Installation
121
Setting the Torch Viewing Position
This procedure adjusts the torch viewing position for the highest signal intensity.
Perform this procedure when:
•
the instrument or software is first installed
•
the instrument is moved to a new location
•
the torch is removed or replaced
•
the plasma induction plates is replaced
1. Switch on the computer and spectrometer, and if required, the autosampler.
2. Switch on the exhaust ventilation system.
3. Start the ICP Winlab software.
4. In the
Options
menu, click on
Autosampler
, and make sure that the software is
configured for the autosampler tray you are using.
5. In the
Tools
menu, click on
Plasma Control
and ignite the plasma. Allow the
torch to warm up for half an hour before you start the alignment.
6. Prepare a solution containing 1 ppm manganese.
7. In the
Tools
menu, click on
Spectrometer Control
.
The Spectrometer Control window appears.
8. Select
Axial
or
Radial
. We recommend that you first do the alignment in one
viewing mode, then select the other viewing mode and repeat the procedure.
9. To view spectra collected during the procedure, open the
Spectra Display
window.
10. Click on
Align View
in the Spectrometer Control window to open the
Align
View
dialog.
11. Select the element and wavelength. This is the wavelength that the system will
use to measure the intensities at different viewing positions and locate the
position that gives the highest intensity. Your options are:
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......