Maintenance
145
GemTip or GemCone
Nebulizer/End Cap
Assembly
Clean with soap and water; rinse thoroughly.
Concentric Glass
(MEINHARD)
Nebulizer
Do
not
use an ultrasonic cleaner. Do
not
use a cleaning
wire. See
Concentric Glass Nebulizer
later in this
chapter.
Mira Mist Nebulizer
For the longest life and best performance, wash your
nebulizer by simply running water as a sample for 10
minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually
unnecessary. Please
DO NOT
wash your nebulizer in
acid or solvents to 'prevent salt build up'.
DO NOT
use
surfactants as they will destroy the surface tension at
the tip and performance.
The tip is very easily
damaged and should NEVER be touched with
fingers, tissues, or anything else
.
Spray Chamber
By handling the spray chamber oil and dirt can
accumulate on it and this will affect performance. To
assure proper wetting wipe the spray chamber with
acetone to remove all oils and dirt.
Pump Head
Remove and clean with water or a mild solvent.
Purge Viewing
Windows/Tubes
Clean with deionized water and dry with a soft cloth.
To clean difficult stains, refer to the information in this
chapter.
Table 5-1.
Recommended Cleaning Procedure
(Continued)
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......