Maintenance
194
5. Check the condition of the tubing piece (Part No. 09985708). If the tubing is
worn, replace it as follows:
•
Obtain a new piece of 1/16-in. O.D. teflon tubing (Part No. 09985708).
•
Using a razor blade, make a straight cut on one end of the tubing.
•
Insert a nut (Part No. 09920546) and ferrule (Part No. 09920518) on the
newly cut end of the tubing with the ferrule oriented as shown in the figure
below.
Figure 5-20
Installing new tubing for the sample inlet. Cross-Flow End Cap,
N0770546.
6. To reassemble the nebulizer make sure that the O-ring is completely seated on
the base of the sample tip, then insert the sample tip into the sample inlet.
4
Nut (Part No. 09920546)
5
1/16 inch O.D. teflon Tubing Piece (Part No. 09985708)
6
Sample Capillary Tubing
Item
Description
1
Ferrule (Part No. 09920515)
2
Nut (Part No. 09920545)
3
1/8 inch O.D. Tubing (Part No. 09985723)
Item
Description
1
2
3
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......