18
Warning labels on the back and side of the instrument
Marking
Location Instructions
for
Safety
Label #1 in
Figure B.
Vmax = 30 V AC, Imax = 0.5 A AC
Vmax = 30 V DC, Imax= 2 A DC
Do not exceed stated voltage and current.
Label #2 in
Figure B.
Air / N
2
For the shear gas, use air or nitrogen.
Label #3 in
Figure B.
Warning - Grounding circuit continuity is vital for the safe operation
of this instrument. Grounding is accomplished by the use of an
IEC309 instrument power plug furnished by the factory and it is not
to be removed during or after installation.
Attention: La continuité de circuit par mettre a la terre est nécessaire pour
l’opération de cet instrument sans risque. L'usine fournit la fiche
d’instrument IEC309 qui mits à la terre l’instrument. N’enlève pas la fiche
pendant ou après l’installation.
Label #4 in
Figure B.
Warning: All field wiring must have insulation suitable for at least 300 V.
Avertissement: Le cablage a pied d’ouvre doit etre certifiés pour un
minimum de 300 V.
Label #5 in
Figure B.
Warning: Disconnect supply before servicing.
Avertissement: Couper l’alimentation avant l’entretien.
.
Air
German
Luft
French
Air
Danish
Luft
Finnish
Ilma
Greek
P
Italian
Aria
Dutch
Lucht
Norwegian
Luft
Portuguese
Ar
Spanish
Aire
Swedish
Luft
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......