Maintenance
219
3. Use the other end of the tubing coupler (Part No. 09920186) to connect the drain
tubing to the peristaltic pump.
Drain tubing with 1 mm i.d. (0.04 in.) can also be used by following these steps:
1. Attach a small piece of 1.5 mm (0.06 in.) i.d. PVC tubing (Part No. 09985729)
to the barbed fitting (Part No. 09920267).
2. Cut the 1 mm i.d. tubing at an angle and insert it into the 1.5 mm tubing.