Maintenance
223
5. Loosen the adjustment screw until the bubbles stop moving.
6. Tighten the adjustment screw
slowly
until the bubbles just start moving again.
7. Tighten the adjustment screw until the bubbles move in a segmented flow (about
1/2 to 1 turn on the peristaltic pump).
8. Turn off the pump. The tension is now correctly set.
Cleaning the Pump
Periodically inspect the pump rollers to make sure they are clean and move freely.
Clean the exterior of the pump with a cloth moistened with water.
The pump head on the peristaltic pump can be removed and cleaned if necessary. A
replacement stainless-steel pump head is available (Part No. N0690360).
Caution
If a spill occurs on the pump, immediately clean it up.
A corrosive spill
could damage the rollers and interior parts of the pump.