Troubleshooting
262
Common Setup Problems:
Glassware bottomed out in torch body (Glassware Holder):Yes/No
Glassware aligned with notch in torch body (Glassware Holder):Yes/No
Copper tape covering igniter spark hole:Yes/No
Igniter finger contacting copper tape:Yes/No
Ground rod located near end of glassware but outside of glasswareYes/No
Fiber Optic located over glassware and is not melted or recessed:Yes/No
Torch position appropriate for sample type:Yes/No
Aqueous = -3, Organics = -3, High Solids = -1
Injector O-Rings not worn, cracked, or cut:Yes/No
Torch Body (Glassware Holder) O-Rings not worn, cracked, or cut:Yes/No
Injector Dry:Yes/No
Spray Chamber draining properly:Yes/No
Pump Tubing clamped:Yes/No
Pump Tubing not worn out:Yes/No
Pump Tubing pumping in proper direction:Yes/No
Radial and Axial windows installed and clean:Yes/No