Error Messages
271
07
Ids 1 out of
level 3
window
The drain to source
current across the
power FET transistor
1 is out of the level 3
Tolerance window.
Contact PerkinElmer Service.
08
Ids 2 out of
level 3
window
The drain to source
current across the
power FET transistor
2 is out of the level 3
Tolerance window.
Contact PerkinElmer Service.
09
Ids 1 higher
than ids
maximum
The drain to source
current across the
power FET transistor
1 is higher than the
maximum allowable
current.
Contact PerkinElmer Service.
10
Ids 2 higher
than ids
maximum
The drain to source
current across the
power FET transistor
2 is higher than the
maximum allowable
current.
Contact PerkinElmer Service.
11
Failed to
reach/
maintain
power at
specified
level
Failed to set power to
a value specified in
the Plasma Controls
Dialog window.
Contact PerkinElmer Service.
12
Failed to
set AUX
flow to
specified
level
Failed to set AUX to
a value specified in
the AUX Setting
Flow Controls
window.
Check argon supply, argon pressure, AUX
flow tubing and connections. If this does not
correct the problem, contact PerkinElmer
Service.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......