Error Messages
274
26
+24V
supply out
of range
(Warning)
This warning
message will appear
if the RF Control
Electronics Supply is
+/-10% out of range.
Contact PerkinElmer Service.
27
DC supply
out of range
(Warning)
This warning
message will appear
if the Instrument
Control Electronics
Supply is +/-10% out
of range.
Contact PerkinElmer Service.
29
Fiber optic
not
detecting
plasma
during
ignition
The firmware expects
the plasma to be on
but the plasma is not
on. The fiber optic
will do a hardware
shutdown.
Check the Optics Interlock. See
Plasma
Ignition and Stability Problems
on page
255for more information. If this does not
correct the problem contact PerkinElmer
Service.
30
Not used
31
Neb
Pressure
out of range
In this hardware
failure the nebulizer
back pressure is out
of the acceptable
range of 0 to 65 psi.
Verify that the NEB Quick-Disconnect fitting
is correctly mated. Examine the nebulizer
tips for blockage. If this does not correct the
problem, contact PerkinElmer Service.
32
Ids 1 is
non-zero
Drain to source
current across the
power FET transistor
1 is a non-zero
current. This
transistor should be
off but current is still
going through it.
Contact PerkinElmer Service.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......