Error Messages
276
40
+DC
Supply out
of range
(Fatal)
This fatal message
will appear if the DC
Supply is 10% out of
range (24V +/-2.4 V).
Contact PerkinElmer Service.
41
Watchdog
Timeout
The microprocessor
malfunctioned and
the instrument lost
functions, but
recovered.
Contact PerkinElmer Service.
42
Fiber optic
detecting
plasma
Hardware error. The
fiber optic is
detecting the plasma
but the plasma is
not
on.
Check the Optics Interlock. See
Plasma
Ignition and Stability Problems
on page
255for more information. If this does not
correct the problem contact PerkinElmer
Service.
43
Ids 1 not 0
Drain to source
current across the
power FET transistor
1 is turned off but
current is still
flowing through this
transistor.
Contact PerkinElmer Service.
44
Ids 2 not 0
Drain to source
current across the
power FET transistor
2 is turned off but
current is still
flowing this
transistor.
Contact PerkinElmer Service.
45
Watchdog
not
working
A hardware/firmware
failure. The safety
timer that protects the
instrument in case the
firmware fails is not
working.
Contact PerkinElmer Service.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......