Error Messages
279
61
Error
reading
post-ignite
power
setting
Parameters for the
ignite command for
the power setting are
invalid.
Contact PerkinElmer Service.
62
Error
reading
post-ignite
AUX flow
Parameters for the
ignite command for
the AUX flow are
invalid.
Contact PerkinElmer Service.
63
Error
reading
post-ignite
neb flow
Parameters for the
ignite command for
the neb flow are
invalid.
Contact PerkinElmer Service.
64
Error
reading
post-ignite
plasma
flow
Parameters for the
ignite command for
the plasma flow are
invalid.
Contact PerkinElmer Service.
65
Learn AUX
failed range
checking
There is a fault in the
pneumatics system.
Contact PerkinElmer Service.
66
Learn
Plasma
failed range
checking
There is a fault in the
pneumatics system.
Contact PerkinElmer Service.
67
Learn AUX
failed slope
checking
The system calculates
a slope to verify that
the AUX flows are
actually increasing.
Verify that the AUX gas is turned on.
Otherwise the plasma will not ignite.
68
Learn
Plasma
failed slope
checking
The system calculates
a slope to verify that
the plasma flows are
actually increasing.
Verify that the plasma is turned on.
Otherwise the plasma will not ignite.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......