Preparing Your Laboratory
53
Laboratory Space Requirements
The Optima 8000 system includes the Optima 8000 instrument, a PolyScience
Recirculating Chiller (or equivalent), a computer, and a printer. The Optima 8000
instrument itself includes optics, electronics, a plasma torch, and an RF power
supply, all housed in a self-contained unit. The minimum door width must be 81 cm
(32 in.).
Optima 8000 Instrument
The Optima 8000 is 132-cm long (52-in.), 76-cm high (30-in.), and 81-cm deep (32-
in.). The bottom of the vent must be a minimum of 75-cm (3-in.) from the top of the
torch compartment chimney. Figure 2-2 illustrates the dimensions of the instrument.
The Optima 8000 weighs 146 kg (321 pounds), or 201 kg (442 pounds) with the
shipping container.
The Optima 8000 may be placed on a bench. The main power cable length is 2.4 m
(8.0 ft.).
To provide access for servicing and space for ventilation, allow at least 61 cm (24
in.) of space between the instrument and the laboratory walls.
Computer and Printer
Refer to the manuals supplied with your computer and printer for dimensions.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......