Preparing Your Laboratory
57
surface mount version) both of which are contained in the N0770425 conduit box kit
that ships with the instrument.
As an alternative you can use the following receptacle part numbers which can be
directly ordered from Hubble. For the US/Canada (20A service) use PerkinElmer
Part Number 09997529 Hubble Part No. C320R6SVL or C320R6W. For Europe
(16A service) use Hubble Part No. C316R6S. Also, Conduit Box-Cast Device Back
Box Hubble Part No. SP 20301 is available for the above Hubble versions.
Computer and Printer
Refer to the guides supplied with your computer and printer for electrical
requirements. The computer and printer must share a common earth ground with the
Optima 8000.
PolyScience Chiller
The specific electrical requirements for the PolyScience
®
WhisperCool
TM
Chiller
are printed on a serial number label located on the back of the unit. The voltage of
the power source must meet the specified voltage
10%. In addition, an adequate
ground connection must be provided.
For 60 Hz installations,
the PolyScience 6106PE V, 60 Hz, 15A unit is supplied
with a 15-ampere, 250-volt Hubbell #4570-C Twist-Lock power plug (NEMA L6-
15P configuration). A Hubbell #4560 or equivalent receptacle (NEMA L6-15R
configuration) is also required, and is supplied with the instrument.
(NEMA: The National Electrical Manufacturer’s Association)
For 50 Hz installations,
to accommodate most countries, the PolyScience 220/240
V, 50 Hz, 15A unit is supplied with two detachable line cords.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......