System Description
68
Wavelength selection is achieved by simultaneous rotation of the prism and grating.
Since the maximum rotation required for either element is not more than ± 2°, the
average wavelength selection time is less than 2 seconds. To further optimize the
analysis time, the wavelengths required are sorted to minimize the change-over time
between any two consecutive wavelengths.
The dual monochromator system enables relatively high slits to be used with no loss
of image quality, which contributes to the high optical throughput. In addition, part
of the slit height is used for the simultaneous measurement of a neon reference
spectrum for wavelength correction.
Two slit width settings are available, optimized for UV and visible wavelengths. The
system automatically selects the correct slit for each analytical measurement.