Installation
97
Connecting the PolyScience Chiller
The PolyScience
®
WhisperCool
TM
Chiller must be used to supply cooling water to
the instrument. The PolyScience
®
WhisperCool
TM
Chiller comes with a complete
instruction manual. You should be familiar with this manual before proceeding.
The PolyScience
®
WhisperCool
TM
is available through PerkinElmer in the
following two configurations:
208/230V, 60 Hz
220/240V, 50 Hz
Connecting the Chiller to the Instrument
Teflon tape (PTFE tape) should be used for all pipe-thread (NPT) fittings.
Use of Teflon tape can prevent leaks.
Connect the instrument to the chiller inlet using coolant drain hose. The direction of
the flow through the system can be controlled by the way the hoses are connected to
the chiller. The "INLET" port will draw liquid into the chiller; the “OUTLET" port
will pump liquid out. See the following figure for the location of the cables.
1. Locate the instrument return/Output coolant line (Part No. N0770342 and
connect to the chiller "INLET" fitting. The instrument water outputs have a
fitting and tubing to connect them to the return or drain. Use 3/8-in. space collars
(Part No. 09920584) at the instrument. The following figure shows the water
supply connections at the instrument.
2. Locate the instrument Input coolant lines (Part No. N0770341, uses 3/8-in
Swagelock fittings) and connect to the chiller “OUTLET” fitting. Use a 3/8-in.
space collars (Part No. 09920584) at the instrument. The following figure shows
the chiller connections. Figure 4-3 on page 101shows the water supply
connections at the instrument.
3. Make sure that the hoses and fittings are tight and that there are no bends or
crimps in the hoses.
Note
A qualified electrician must install the single wall receptacle for the
PolyScience
®
WhisperCool
TM
chiller.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......