Installation
100
Remote Chiller Operations
The remote chiller will automatically be on when the Optima 8000 Spectrometer is
warming up. The warming up period includes just being turned on, coming out of
Sleep mode or Standby mode.
If the instrument has just been turned on, starting WinLab32 will automatically begin
warm up. Initiating WinLab 32 will have no effect on the chiller if the chiller is
already on.
If the Optima 8000 Spectrometer is in Sleep or Standby mode and the instrument is
still ignited, the chiller will remain on. If the instrument is not ignited in Sleep or
Standby mode the chiller will turn itself off.
5
Coolant Drain Hose (Part No. N0770342) White Tie Wrap
6
From Optima Out
7
Coolant In
8
To Optima In
9
Coolant Supply Hose (Part No. N0770341) Red Tie Wrap
10
Line Cord Plug
11
Chiller
Note
The plasma should
never
be ignited if the chiller is off.
Item
Description
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......