Permobil C400 Corpus 3G Lowrider
Troubleshooting R-net
Joystick Error
The commonest cause for this error is that the joystick was moved away from its central position before and during the
time at which the control system was switched on. The screen for a shifted joystick is displayed for 5 seconds. If the
joystick is not released during this time, a joystick error is registered. Even if an error screen is not displayed, the error
and the number times it arises is registered in the system log.
• Ensure that the joystick is in the central position and start up the control system.
If the error persists, the joystick or joystick module may be defective. Read more in
Repairing defective units
, Page
120.
Low Battery
This occurs when the control system detects that the battery voltage is lower than 16 V.
• Check the batteries and their connection to the control system.
If the error persists after the batteries and connections have been checked, the power module may be defective. Read
more in
Repairing defective units
, Page 120.
High Battery
This occurs when the control system detects that the battery voltage is higher than 35 V. The commonest causes for
this error are that the battery has been overcharged or a poor connection between the control system and the
batteries.
• Check the batteries and their connection to the control system.
If the error persists after the batteries and connections have been checked, the power module may be defective. Read
more in
Repairing defective units
, Page 120.
Brake Error
This occurs when the control system detects a problem in the solenoid brakes or the connections to them.
1505 - M1 Brake Error
1506 - M2 Brake Error
• Check the solenoid brakes, their cables and the connections to the control system.
If the error persists after the checks listed above, the power module may be defective. Read more in
Repairing defec-
tive units
.
Motor Error
This occurs when the control system detects that a motor has been disconnected.
3B00 - M1 Motor Error
3C00 - M2 Motor Error
• Check the motors, their cables and the connections to the control system.
If the error persists after the checks listed above, the power module may be defective. Read more in
Repairing defec-
tive units
, Page 120.
Inhibit Active
This occurs when one of the inhibit signals is active and is in blocked mode.
The last two digits of the error code indicate the active inhibit signal. The code is hexadecimal.
1E01 - For inhibit signal 1.
1E09 - For inhibit signal 9.
1E0A - For inhibit signal 10.
• Cycle the voltage. This will deactivate the block mode, which may remedy the error.
• Check all connections and switches for the indicated inhibit signals.
Joystick Calibration Error
This occurs when joystick calibration has been unsuccessful.
• Go to OBP mode and recalibrate.
If the error persists, the joystick module may be defective. Read more in
Repairing defective units
, Page 120.
116
Summary of Contents for C400
Page 1: ...Service manual C400 Corpus 3G Lowrider ...
Page 2: ......
Page 4: ...Text ...
Page 5: ......
Page 8: ......
Page 10: ...Permobil C400 Corpus 3G Lowrider Introduction 10 ...
Page 12: ...Permobil C400 Corpus 3G Lowrider Serial Number labels 12 ...
Page 94: ...Permobil C400 Corpus 3G Lowrider Repairs 94 ...
Page 110: ...Permobil C400 Corpus 3G Lowrider Control system 110 ...
Page 123: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview CABLING OWERVIEW Chassis 124 Seat 126 123 ...
Page 124: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Chassis 124 ...
Page 125: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 125 ...
Page 126: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Seat 126 ...
Page 127: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 127 ...
Page 128: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 128 ...
Page 131: ......
Page 132: ...A 205334 UK 0 www permobil com ...