Permobil C400 Corpus 3G Lowrider
Repairs
Assembly
Assemble in the reverse order.
1.
Figure 90. The seat is held in place by four screws.
Mount the seat using the four screws. See fig. 90. It
can be mounted in three different positions, depend-
ing on which chassis it is being mounted on and the
depth of the seat. See the table below. Use a torque
wrench to tighten the screws.
Tightening torque 24
Nm.
Mounting position
Seat depth
C400
C500
C500S
370 mm./15"
1
1
3
395 mm./16"
1
1
3
420 mm./17"
1
1
3
445 mm./18"
1
1
3
470 mm./19"
1
1
3
495 mm./20"
1
1
3
520 mm./21"
2
2
3
545 mm./22"
3
3
3
570 mm./23"
3
3
3
2.
Figure 91. Connectors on the control panel cabling (1)
and seat tilt cabling (2).
Reconnect the control panel cabling (1) and on seat
equipped with seat tilt, the seat tilt cabling (2) at the
contacts on the cabling. The contacts are at the right
hand side of the seat angle mechanism, next to the
other cabling. See fig. 91.
62
Summary of Contents for C400
Page 1: ...Service manual C400 Corpus 3G Lowrider ...
Page 2: ......
Page 4: ...Text ...
Page 5: ......
Page 8: ......
Page 10: ...Permobil C400 Corpus 3G Lowrider Introduction 10 ...
Page 12: ...Permobil C400 Corpus 3G Lowrider Serial Number labels 12 ...
Page 94: ...Permobil C400 Corpus 3G Lowrider Repairs 94 ...
Page 110: ...Permobil C400 Corpus 3G Lowrider Control system 110 ...
Page 123: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview CABLING OWERVIEW Chassis 124 Seat 126 123 ...
Page 124: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Chassis 124 ...
Page 125: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 125 ...
Page 126: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Seat 126 ...
Page 127: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 127 ...
Page 128: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 128 ...
Page 131: ......
Page 132: ...A 205334 UK 0 www permobil com ...