Main circuit breaker
Figure 193. Main circuit breaker.
The main circuit breaker also functions as a battery isolator although
it is referred to as the main circuit breaker in the user’s manual.
The main circuit breaker is automatic and can be reset when it has
been triggered. Reset it by putting the switch to the On position.
NOTICE
Investigate tripped main circuit breakers
A tripped main circuit breaker often indicates a major electrical fault. The
cause should be carefully investigated before resetting.
NOTICE
Before using the main circuit breaker
Always switch off the power to the control panel before switching off the
main power with the main circuit breaker.
ON / OFF
ON / OFF
243