Reassembly
1. Follow the disassembly steps in reverse.
2. Update the diaphragm pump maintenance counter from the menu
[Maintenance] [Information]
[Backing Pump] [Reset timer]
(see chapter “Information” of the operating manual).
Diaphragm replacement
► See the diaphragm pump Operating Instructions (see chapter “Applicable Documents”).
5.10 Solenoid valve replacement
Disassembly
1. Create an air inlet on the analyzer cell from the menu
[Maintenance] [Maintenance turbo pump
& cell] [Stop and vent]
(see chapter “Maintenance turbo pump & cell” in the operating manual).
2. Power off the leak detector (see chapter “Powering Off”).
3. Remove the cover (see chapter “Cover disassembly/reassembly”).
Service - Maintenance
21/30