Reassembly
3
F1
2
1
4
1
F1 Filament 1
3
Connectors
1
Fixing screw and washer
4
Centering stops
2
Filament strip
The connectors are perpendicular to the supporting surface and parallel to each other.
2
5
3
4
1. Place the new filament on its housing against both centering stops. Fix it with the screw and
washer provided with the filament.
2. Fit the fastening clips on the connectors on each side of the filament.
3. Reassemble the analyzer cell (see chapter “Disassembly/reassembly of the analyzer cell”).
4. Update the replaced filament maintenance counter from the menu (see chapter “Information” of
the operating instructions).
5.6 Fan maintenance
5.6.1 Air filter replacement
► Power off the leak detector (see chapter “Powering Off”).
Service - Maintenance