12
Installation
5
Installation
5.1
Assembly
5.1.1
Rack installation
The electronic drive unit housing TCP 1200 consists of a 19" component rack for instal-
lation in a rack cabinet.
Insert the device horizontally in the guide rails of a 19"/3HE rack and screw on the
front panel securely with four fixing screws.
Ensure sufficient air circulation if installed in enclosed housings.
Do not hinder device internal air flow and removal.
Ensure permissible ambient conditions in the rack cabinet
.
5.2
Connection "external pump"
The external connection of a Pfeiffer Vacuum turbopump to the electronic drive unit TCP
1200 is possible, using shielded connection cables of different lengths from the Pfeiffer
Vacuum accessories programme.
The cable for control and operation of the turbopump is uniquely designated by plugs and
sockets.
●
HAN-Q5 plug and RJ45 plug for the connection to the TCP 1200
●
HAN-Q5 socket and M12 socket for the connection to the turbopump
NOTICE
Ensure free convection
Incorrect installation can cause damage by over-heating or create a fire hazard.
Maintain a minimum distance of 50 mm from cooling vents to adjacent components
or boundaries.
Install the device upright.
WARNING
Danger of electric shock
The system is only free of voltages if the mains plug is disconnected.
Switch off the main switch and disconnect the mains plug before all work.
Secure against unintentional restarting.
Ensure the complete stop of the pump (f < 1).
CAUTION
Risk of tripping!
Risk of tripping when working in the installation area.
Place supply lines in such a way that no tripping hazards occur.