Connection "remote"
15
6
Connection "
remote
"
Remote control is possible via the 26-pin D-sub connector labelled "
remote
" on the elec-
tronic drive unit. The accessible individual functions are mapped to "PLC levels".
Remove the remote plug from the TCP 1200 and connect a remote control unit. Pin
assignment of the connector according to table.
Shielded connectors and cables must be used.
The following information display the factory setting. Configuration is possible
using the Pfeiffer Vacuum parameter set.
6.1
Pin assignment
6.2
Operation via "
remote
" connection
6.2.1
+24 V DC* Output / Pin 1
Inputs 2 - 6 and the connections to Pins 10, 13, 14 are activated by connecting them with
+24 V DC to Pin 1 (active high). They can also be activated via an external PLC. The
functions are deactivated by "PLC high level" and by "PLC low level".
●
PLC high level: +13 V to +33 V
●
PLC low level: -33 V to +7 V
●
Ri: 7 k
Ω
●
I
max
< 210 mA (with RS-485, if existing)
1
9
10
18
19
26
Pin Function
Designation factory settings
1
+24 V DC output (V+)
Reference voltage for all digital in- and outputs
2
DI1
Enable venting; open: no; V+: yes
3
DI Motor pump
Drive motor; open: off; V+: on
4
DI Pumping station
Open: off; V+: on and error acknowledgement
5
DI Standby
Standby rotation speed; open: off; V+: on
6
DI2
Heating; open: off; V+: on
7
AI+ Rotation speed setting mode Set value in rotation speed setting mode;
2-10 V DC = 20-100% of the nominal rotation speed
8
DO1
Rotation speed switch point attained;
GND:no; V+: yes (I
max
= 50 mA/24 V)
9
DO2
GND: error; V+: no error (I
max
= 50 mA/24 V)
10
DI3
Sealing gas; open: off; V+: on
11
AI- Rotation speed setting mode
GND
Set value in rotation speed setting mode; GND
12
AO1
Actual rotation speed;
0-10 V DC is equivalent to 0-100%; R
L
> 10 k
Ω
13
DI Error acknowledgement
Error acknowledgement: V+ pulse (min 500 ms)
14
DI Remote priority
Control via interface "remote"; open: off
V+: set and priority over other digital inputs
15
Relais 1
Connection to Pin 16 if relay 1 is inactive
16
Relais 1
Rotation speed switchpoint attained;
relay contact 1 (U
max
= 50 V DC; I
max
= 1 A)
17
Relais 1
Connection to Pin 16 if relay 1 is active
18
Relais 2
Connection to Pin 19 if relay 2 is inactive
19
Relais 2
No error; relay contact 2 (U
max
= 50 V DC; I
max
= 1 A)
20
Relais 2
Connection to Pin 19 if relay 2 is active
21
Relais 3
Connection to Pin 22 if relay 3 is inactive
22
Relais 3
Warning; relay contact 3 (U
max
= 50 V DC; I
max
= 1 A)
23
DO Remote priority
GND: off; V+: remote priority active
24
RS-485 D+
according to specifications and Pfeiffer Vacuum protocol
25
RS-485 D-
according to specifications and Pfeiffer Vacuum protocol
26
Ground (GND)
Reference ground for all digital inputs and all outputs