The Pfeiffer Vacuum parameter set
23
9.2.4
Status requests
046 Cfg Rel R2
Configuration Relay 2
Options see
[P:019]
7 RW
0
21
1
x
047 Cfg Rel R3
Configuration Relay 3
Options see
[P:019]
7 RW
0
21
3
x
050 SealingGas
Sealing gas
0 = off
1 = on
0 RW
0
1
0
x
055 Cfg AO1
Configuration output AO1
0 = Actual rotation speed
1 = Power
2 = Current
3 = always 0 V
4 = always 10 V
5 = follows AI1
6 = Pressure value 1
7 = Pressure value 2
8 = Control fore-vacuum
7 RW
0
8
0
x
057 Cfg AI1
Configuration input AI1
0 = Disconnected
1 = Set value rot. speed setting mode
7 RW
0
1
0
x
060 CtrlViaInt
Control via interface
1 = Remote
2 = RS-485
4 = PV.can
8 = Field bus
16 = E74
255 = Unlock interface selection
7 RW
1
255
1
x
061 IntSelLckd
Interface selection locked
0 = off
1 = on
0 RW
0
1
0
x
062 Cfg DI1
Configuration input DI1
Settings
[P:063/064]
0 = deactivated
1 = Enable venting
2 = Heating
3 = Sealing gas
4 = Run-up time control
5 = Rotation speed setting mode
7 = Enable HV sensor
7 RW
0
7
1
x
063 Cfg DI2
Configuration input DI2
Options see
[P:062]
Settings
[P:062/064]
7 RW
0
7
2
x
064 Cfg DI3
Configuration input DI3
Options see
[P:062]
Settings
[P:062/063]
7 RW
0
7
3
x
#
Display
Designation
Functions
Da
ta
t
ype
Ac
ce
ss
Unit
min max
default
#
Display
Designation
Functions
Dat
a
ty
pe
A
cce
ss
Unit
min max
default
300 RemotePrio
Remote priority
0 = no
1 = yes
0 R
0
1
301 OpFluidDef
Operating fluid deficiancy
0 = no
1 = yes
0 R
0
1
302 SpdSwPtAtt
Rotation speed switchpoint attained
0 = no
1 = yes
0 R
0
1
303 Error code
Error code
4 R
304 OvTempElec
Excess temperature electronic drive unit
0 = no
1 = yes
0 R
0
1
305 OvTempPump
Excess temperature pump
0 = no
1 = yes
0 R
0
1
306 SetSpdAtt
Set rotation speed attained
0 = no
1 = yes
0 R
0
1
307 PumpAccel
Pump accelerates
0 = no
1 = yes
0 R
0
1
308 SetRotSpd
Set rotation speed (Hz)
1 R
Hz
0
999999
309 ActualSpd
Active rotation speed (Hz)
1 R
Hz
0
999999
310 DrvCurrent
Drive current
2 R
A
0
9999.99
311 OpHrsPump
Operating hours pump
1 R
h
0
65535
x
312 Fw version
Firmware version electronic drive unit
4 R
313 DrvVoltage
Drive voltage
2 R
V
0
9999.99
314 OpHrsElec
Operating hours electronic drive unit
1 R
h
0
65535
x
315 Nominal Spd
Nominal rotation speed (Hz)
1 R
Hz
0
999999
316 DrvPower
Drive power
1 R
W
0
999999
319 PumpCycles
Pump cycles
1 R
0
65535
x
324 TempPwrStg
Temperature power stage
1 R
°C
0
999999
326 TempElec
Temperature electronic
1 R
°C
0
999999
336 AccelDecel
Acceleration / Deceleration
1 R
rpm/s
0
999999