26
The Pfeiffer Vacuum parameter set
9.3.4
Analog output on "
remote
"
Configuration via parameter
[P:055]
.
9.3.5
Analog input on "
remote
"
Configuration via parameter
[P:057]
.
9.3.6
Control via interface
Configuration via parameters
[P:060]
and
[P:061]
.
9.4
Operation with the Pfeiffer Vacuum parameter set
9.4.1
Factory settings
The electronic drive unit is pre-programmed in the factory. This guarantees proper, reli-
able turbopump operation without the need for additional configuration.
9.4.2
Checking the adjustments
Before operating with parameters, check set values and control commands for their
suitability for the pumping process.
Remove the remote plug from electronic drive unit if required.
Option
Description
0 = Rotation speed
Rotation speed signal; 0 - 10 V DC = 0 - 100 % x f
Nominal
1 = Power
Power signal; 0 - 10 V DC = 0 - 100 % x P
max
2 = Current
Current signal; 0 - 10 V DC = 0 - 100 % x I
max
3 = always 0 V
Always GND
4 = always 10 V
Output of continously 10 V DC
5 = follows AI1
Follows the analog input 1
6 = Pressure value 1
Pressure value signal;
0 V: error
1 V: underrange
1.5 - 8.5 V: p (hPa) = 10
(U-5.5 V)
9 V: overrange
7 = Pressure value 2
8 = Control fore-vacuum
Fore-vacuum signal; control of Pfeiffer Vacuum
turbo pumping stations
Option
Description
0 = Switched off
Connection is deactivated
1 = Set value in rotation speed setting
mode
Rotation speed setting mode via pin 7 (0 - 10 V) and pin 11
(GND)
Option [P:060]
Description
1 = remote
Operation via connection "
remote
"
2 = RS-485
Operation via connection "
RS-485
"
4 = PV.can
For service purposes only
8 = Field bus
Operation via field bus
16 = E74
Operation via connection "
E74
"
Option [P:061]
Description
0 = off
Interface selection via [P:060]
1 = on
Interface selection locked