36
Operation
●
Edit mode for the parameter is ended. The arrow ( –––
►
) disappears.
Edit mode( –––
►
) will disappear again automatically under the following condi-
tions and without accepting the value to be changed:
●
Input disruption or no key operation for more than 10s.
●
If an error occurs.
●
Key "Pumping station ON/OFF" has been pressed.
●
If line 3 = empty , "data not changed" will be displayed.
10.7 Pressure measurement
Connection and use of a high vacuum gauge enables pressure measurement in the vac-
uum system via the TCP.
10.7.1 Displaying the gauge type
Parameter
[P:794]
= 1 (display extended parameter set)
Select or enter parameter
[P:738] Gaugetype
.
Gauges with identical surge impedance are only recognized as a group (e.g. CMR?).
Manual input of the exact gauge type is also possible via the parameter [P:738].
10.7.2 Displaying the active pressure value
Parameter
[P:794]
= 1 (display extended parameter set)
Select parameter
[P:340] Pressure
.
10.8 Switching off
Press key
on the DCU front panel.
Pressure measurement with the DCU
An exact pressure measurement is not possible with the DCU. This is particularly true
with linear gauges in the lower pressure range.
Use suitable measuring instrument.
Display example
Meaning
TPR 2xx
Gauge TPR 280 connected
CMR ?
Gauge of the CMR Group connected, exact type not yet set
noGaug
No pressure gauge connected
Display example
Meaning
–––––– mbar
No pressure gauge connected
< 5E-4mbar
Measuring range not reached (dependent on the device)
> 1E3mbar
Measuring range exceeded (dependent on the device)
6.3E-9mbar
Valid pressure measurement
id fam mbar
Gauge type not yet identified; see [P:340]
Error
Error in pressure gauge