Safety
7
2.2
Proper use
●
The external electronic drive unit TCP 1200 only serves the operation of
Pfeiffer Vacuum turbopumps of the series HiPace®:
– HiPace 1200 with TCP 1200
– HiPace 1500 with TCP 1200
– HiPace 1800 with TCP 1200
– HiPace 2300 with TCP 1200
●
The external electronic drive unit TCP 1200 serves as voltage supply, control and
monitoring of a turbopump and connected peripherals.
Only operate the pump with the respective electronic drive unit. The use of other than
the intended electronic drive units is not permitted.
2.3
Improper use
Improper use will cause all claims for liability and warranties to be forfeited. Improper use
is defined as usage for purposes deviating from those mentioned above, especially:
●
connection to pumps or units which are not suitable for this purpose according to their
operating instructions
●
connection to units which have exposed voltage-carrying parts
●
connection to mains supplies, which do not confirm to the regulations IEC 61010 or
IEC 60950
●
operation of the devices in areas with ionizing radiation
NOTICE
EC conformity
The manufacturer's declaration of conformity becomes invalid if the operator modifies
the original product or installs additional components.
Following installation into a plant and before commissioning, the operator must check
the entire system for compliance with the valid EU directives and reassess it accord-
ingly.
warranty seal
Closure seal
The product is sealed at the factory. Damaging or removal of a closure seal leads to the
loss of liability and warranty entitlements.
Do not open the product within its warranty period!
For process-related shorter maintenance intervals please contact the Pfeiffer Vacu-
um Service.