5.1.4 Using assembling trestle
2
1
3
Fig. 11:
Analyzer on the assembling trestle
1 Analyzer
3
Assembling trestle
2 Support
Procedure
1. Clamp the assembling trestle onto a stable table.
2. Place the analyzer with the support on the assembling trestle.
5.1.5 Removing transportation protection
NOTICE
Impairment from contamination and damage
Touching the devices or components with bare hands increases the desorption rate and leads to in-
correct measurements. Dirt (e.g. dust, fingerprints, etc.) and damage impair the function.
► During assembly and maintenance work on high or ultra high vacuum systems, always wear
clean, lint-free and powder-free laboratory gloves.
► Only use clean tools.
► Make sure that the connection flanges are free of grease.
► Remove protective caps and protective covers from flanges and connections only when necessa-
ry.
► Remove the analyzer’s transport protection only when necessary.
► Carry out all work in a well lit area.
Installation
28/52