Mixing pump RITMO L / RITMO M Overview - Operation
Maintenance
2016-02-17
63
48.1 Blow dry water fitting
Fig. 108: Connect and fix the air
1. Connect air hose with Geka-coupling and EWO-coupling at the
compressed air connection (1) and to the water connection (2).
48.2 Blow dry water fitting RITMO L
Fig. 109: Blow dry water fitting
1. Insert the connector air compressor (1) into the blue socket
outlet.
2. Switch main switch (2) clockwise to position “I”.
3. Keep water supply button (3) pressed for about 15 seconds.
4. The water is blown out of the fitting using compressed air.
5. Open all water valves and blow once again with compressed
air.
6. Remove the plug (1).
7. Turn the main switch (2) to position “0”.
48.3 Blow dry water fitting RITMO M
Fig. 110: Blow dry water fitting
1. Insert the connector Air compressor into the blue socket outlet.
2. Press the green push button (1) operation “ON”.
3. Keep water supply button (2) pressed for about 15 seconds.
4. The water is blown out of the fitting using compressed air.
5. Open all water valves and blow once again with compressed
air.
6. Remove connector air compressor.
7. Press the red push button (3) operation “OFF”.
49 Maintenance
49.1 Safety
Personnel
The maintenance works described here can be carried out by
the operator, unless marked otherwise.
Some maintenance work must only be carried out by specially
trained technical personnel or exclusively by the manufacturer.
Work on the electrical system must, in principle, be carried out
only by electricians.