background image

pH ELECTRODE

Model : PE-03

Your purchase of this pH

ELECTRODE marks a step

forward for you into the

f i e l d   o f   p r e c i s i o n

measurement. Although

this pH ELECTRODE is a

complex and delicate

instrument, its durable

structure will allow many

years of use if proper

operating techniques are

developed. Please read the
following instructions
carefully and always keep
this manual within easy
reach.

       OPERATION MANUAL

Summary of Contents for PE-03

Page 1: ...eld of precision measurement Although this pH ELECTRODE is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read...

Page 2: ...TABLE OF CONTENTS 1 SPECIFICATIONS 1 2 HELPFUL OPERATING TECHNIQUES 2 3 CALIBRATION AND MEASUREMENTS 3 4 ELECTRODE STORAGE 4 5 ELECTRODE CLEANING 4 6 RECONDITIONING 5...

Page 3: ...otential 7 1 pH for pH Value Body Material Epoxy Connector BNC Mechanical With protection bottle on the electrode Protection head Dimensions Body length 160 mm Body Dia 12 mm Cable length 1000 mm Appl...

Page 4: ...er 3 Vigorously stir the electrode in the sample buffer or rinse solution This action will bring solution to the electrode s surface more quickly and improve speed of response 4 After exposure to a sa...

Page 5: ...le if the span is 10 or more in error a reading of 4 3 or higher for this example the electrode should be cleaned and retested see the Electrode Cleaning Section or reconditioned see the Reconditionin...

Page 6: ...ectrode then the o ring then insert the electrode into the bottle and firmly tighten the cap If the solution in the soaker bottle is missing fill the bottle with pH 4 buffer 5 ELECTRODE CLEANING Coati...

Page 7: ...rse the electrode tip in 0 1N HCL for 15 seconds rinse in tap water and then immerse tip in 0 1N NaOH for is seconds and rinse in tap water Repeat this sequence three times and then recheck the electr...

Reviews: