background image

33

32

CLEAN THE NAVIGATION CAMERA

CAUTION: If the lens is soiled, DO NOT touch the lens.

NOTES 

1. Remove debris from the navigation camera using a facility vacuum source.

2. Wipe the surface of the navigation camera body and cable with a soft cloth dampened with water.

3. Dip a cotton swab into a lens cleaning fluid and softly wipe the lens. Make sure all the debris is removed from 

the lens surface.

4. When the cleaning fluid has evaporated, use a lens cleaning tissue to softly wipe the lens. Wipe from the center 

of the lens to the edge of the lens in a counterclockwise direction.

PERIODIC MAINTENANCE

CLEANING

!

Warnings

Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO 

NOT use any component if damage is apparent.

Perform recommended maintenance as indicated in these instructions. Only trained and experienced healthcare 

professionals should maintain this equipment.

DO NOT service this equipment. If you require service, contact your PHACON sales representative.

NOTE: If any component must be discarded and replaced, see the 

Disposal/ Recycle

 section and the Accessory 

Information section.

CLEAN THE HEAD

NOTES

• Make sure the Patient has been removed from the head and the head has been removed from the tray.

• Thoroughly remove all debris from the Patient lever to make sure the lever has complete freedom of movement.

Failure to comply may prevent the proper installation of the temporal bone insert into the bone cavity

during set up.

1. Remove loose debris from the surface of the head, including the mounting plate, cables, and suction port

using an external vacuum source.

2 . Carefully remove all debris from the risk structure contact pins of the head using a dry brush and an external

vacuum source.

3. While articulating the Patient lever, remove all debris from around the lever using a dry brush and an

external vacuum source. Make sure the lever has complete freedom of movement.

4 . Remove all debris from the spillways within the head using a dry brush and an external vacuum source.

Make sure the spillways are clear to allow water to low freely during the use of irrigation.

5 . Wipe all the surfaces of the head, including the contact pins, insert lever, and spillways with a soft cloth wet

with water.

See the 

Cleaning

 section.

Lens is covered by debris.

Navigation Camera with

Thumb Screw

Replace the camera or thumb screw.

See the 

Accessory Information 

section.

Surface is cracked or broken. Lens is

scratched. USB cable or USB exten-

sion cable insulation is cut or frayed. 

Thumb screw is damaged or missing.

See the 

Cleaning 

section.

Contact pins are covered with debris.

T-Bone insert lever is soiled and/or 

does not move freely.

Before each use

Laptop Computer

See the 

instructions for use supplied 

with the laptop computer

.

Replace the laptop computer. See the

Accessory Information 

section.

INTERVAL

COMPONENT

ACTION

CRITERIA

The laptop computer is operating

improperly or is damaged.

Head

Replace the module. See the 

Accessory

Information 

section.

Model surface, suction port, or

mounting plate is cracked or broken.

Contact pins are bent or broken.

T-Bone insert lever is broken. USB 

cable or instrument ground cable in-

sulation is cut or frayed. Ground

cable with magnet or hook-and-loop 

tape is damaged.

Head Holder with Tray

Replace the head holder. See the

Accessory Information

 section.

Holder surface or ball-joint mounting

plate is cracked or broken. Lever(s) 

or lock(s) are broken or damaged.

See the 

Cleaning

 section.

Ball-joint mounting plate is soiled 

and/or does not move freely.

Lever(s) or lock(s) are soiled and/or 

do not move freely.

Pointer

Replace the pointer. See the 

Accessory

Information

 section.

Surface or tip is cracked or broken.

Before each use

Manual Instrument

Tracker with Thumb Screw

Replace the tracker. See the 

Accessory

Information

 section.

Surface is cracked or broken. Thumb

screw is damaged or missing.

Drill Tracker with O-ring

Replace the tracker or O-ring. See the

Accessory Information

 section.

Surface is cracked or broken. O-ring 

is damaged or missing.

Electrical Bur Adapter

Replace the electrical bur adapter. See 

the 

Accessory Information

 section.

Adapter is damaged or missing.

Summary of Contents for S-00024

Page 1: ...USER MANUAL PHACON TEMPORAL BONE TRAINER BILATERAL...

Page 2: ...DISCOVER THE INTRODUCTION INTO THE PHACON TRAINER your laptop with navigation software your tracking camera your holder tray with head your trackers your patient your pointer...

Page 3: ...FORMATION 8 TRAINER COMPONENTS 9 INSTRUCTIONS 11 LAPTOP DISPLAY SCREEN 12 BUTTONS 13 OPERATE THE TRAINER 15 Enter the application 18 Perform setup for navigation 19 Perform patient calibration 20 Perf...

Page 4: ...rument tracker set for ENT surgery RE 00005 DESCRIPTION REF 3 instrument linkage cable RE 00009 1 notebook with PHACON simulation software RE 00011 1 USB 3 0 hub ZS 00006 1 computer mouse RE 00012 1 p...

Page 5: ...d lever lock to fasten the head to the ball joint mounting plate 3 Rotate the head to the desired position Rotate the ball joint lever lock to secure the head in place 4 Connect the camera USB extensi...

Page 6: ...pecific panel type C Camera Line of Sight Used to display the image within the navigation camera s line of sight D Simulation Time Used to display the start time of the simulation the current time and...

Page 7: ...e software German English Reset the default settings for brightness contrast and slice position in the 4 windowpane in the simulation mode Close and exit the program End the SIMULATION or training ses...

Page 8: ...click on the PHACON Trainer icon located on the laptop computer desktop to log into the training program ENTER THE APPLICATION 2 Type in your user name and password If you do not have an account you...

Page 9: ...fields are optional 4 From the MAIN MENU screen click the SIMULATION button Fig 3 Fig 3 Click on the SETUP button to perform instrument and patient calibration for optimal navigation results See To P...

Page 10: ...of the pointer drill or manual instrument The five calibration points are visible on the Temporal Bone Patient and within the 3 D windowpane during calibration Sequence numbers and a progress bar are...

Page 11: ...atient Make sure the tracking shapes are facing the camera and are within the camera line of sight 3 After visual and audible feedback indicates successful data acq uisition change the angle of the to...

Page 12: ...middle mouse button may also be used When using navigation the CT image slice is selected automatically based on the tip location of the pointer drill or manual instrument To view different slices of...

Page 13: ...he 3 D windowpane Fig 11 ADJUST THE OPACITY VALUE OF AN ANATOMICAL STRUCTURE The opacity of a specific anatomical structure may be changed to provide better visualization within and behind the Patient...

Page 14: ...ING which indicates that the recording has been started 2 To end the recording clickon the button RECORDING 3 To save the recorded video click on the CLASSIC but ton to change to the simulation mode N...

Page 15: ...ND button to end the training session and return to the MAIN MENU screen END THE SIMULATION OR TRAINING SESSION VIEW THE TRAINING SESSION RESULTS 1 From the MAIN MENU screen click on the RESU LTS butt...

Page 16: ...r group results Fig 20 1 Click on the ADD GRAPH button to compare your results with other group results Fig 20 2 From the COMPARISON WITH OTHER GROUPS dialog box clickon the GROUP drop down list and s...

Page 17: ...thin the head using a dry brush and an external vacuum source Make sure the spillways are clear to allow water to low freely during the use of irrigation 5 Wipe all the surfaces of the head including...

Page 18: ...s in the left column of the NVIDIA Control Panel 3 In the Global Settings klick on High performance NVIDIA processor 4 If the Apply button does not appear please select High performance NVIDIA process...

Page 19: ...e head in the free space on the top of the holder tray Now you can close the case lid top gently TRAINER STORAGE HOW TO PACK THE TRANSPORT CASE MESSAGE CAUSE ACTION The LOGIN password or user name is...

Page 20: ...ed with accessories not manufactured by PHACON or if repairs are performed by any party other than authorized PHACON repair personnel Except as set forth above phacon makes no warranties with respect...

Page 21: ...Germany tel 49 0 3 41 47 83 97 32 fax 49 0 3 41 47 83 91 74 www phacon de email info phacon us 4470 Chamblee Dunwoody Rd Suite 463 Dunwoody GA 30338 USA tel 1 833 2 PHACON 2 742266 fax 470 243 0494 w...

Reviews: