background image

35

34

CLEAN THE HEAD HOLDER WITH TRAY

NOTES

1. Remove loose debris from the exterior of the tray using an external vacuum source.

2 . While articulating the ball-joint mount and each lever, remove all debris from the ball-joint mount using a

dry brush and an external vacuum source. Make sure the ball-joint mount and levers have complete

freedom of movement.

3. Wipe all the surfaces of the tray, including the ball-joint mount and each lever, with a soft cloth wet with

water.

CLEAN THE TRAINER ACCESSORIES

NOTES

1. Wipe all the trainer accessories, including the pointer, trackers, cables and electrical bur adapter with a

soft cloth wet with water.

2. After all the trainer components and accessories have been cleaned properly, see the Storage and Handling

section.

!

Warnings

: ALWAYS disconnect the laptop computer from the facility power source before cleaning. See the instructions for use supplied with 

the laptop computer. ALWAYS disconnect and remove all system components from the laptop computer before cleaning system components.

Cautions:

 ALWAYS clean the surgical instruments after a training session. See the instructions for use and/or care instructions supplied with 

the surgical instrument. DO NOT sterilize this training system or any component of the system. DO NOT use solvents, lubricants, or other chem-

icals, unless otherwise specified. DO NOT allow a liquid to run directly into any electrical or mechanical connection or opening. Moisture may 

cause corrosion to electrical and/or mechanical components. DO NOT bend any connector pins during cleaning.

The ground cable with magnet is not in-

stalled on the manual instrument body or 

is not making contact with the manual in-

strument body.

Make sure the ground cable with magnet

is installed properly.

See the 

To Install the Temporal Bone Insert

and Connect Surgical Instruments

 section.

The trainer is not detecting critical risk

structures.

PROBLEM

CAUSE

ACTION

TROUBLESHOOTING

Choose right graphic adapter NVIDIA

1: Open the selection window (context menu) with a right click on your    

     desktop and choose NVIDIA Control Panel.

2: Select the “Manage 3D settings” in the left column of the NVIDIA  

     Control Panel.

3: In the Global Settings klick on “High-performance NVIDIA processor”.

4: If the “Apply” button does not appear, please select  “High-performance 

     NVIDIA processor” again.

5: Save settings by click on the “apply” button.

5.

4.

3.

2.

The navigation system is not functioning 

or is not functioning properly.

The software application locks up.

The application is experiencing an inter-

nal malfunction.

Press the CTRL/ALT/Delete buttons simultane-

ously to perform a soft reboot. Press the ON/

OFF button to perform a hard reboot.

PROBLEM

CAUSE

ACTION

TROUBLESHOOTING

The camera lens is covered with debris.

Clean the camera lens. See the 

To Clean the

Navigation Camera

 section.

An unobstructed lineof- sight does not 

exist between the navigation camera lens 

and the tracker.

Make sure the lineof–sight between the

navigation camera lens and the tracker is

unobstructed.

The navigation system is not accurate 

based on visual indications.

The pointer, drill, or instrument is out of

calibration.

Recalibrate the pointer, drill, or instrument. 

See the 

Perform Tool Calibration

 section.

During scrolling through the CT-images,

always the same area is shown.

The view mode ENDOSCOPIC is

enabled.

Change to the Classic mode in the view

panel options.

The virtual 3-D model has a delayed

reaction.

The notebook is accu driven. The power

supply is not connected.

Connect the power supply.

An instrument tracker is in the line-of-sight 

of the camera in the classic mode.

Remove the tracker from the line-of-sight

of the camera.

The electrical bur adapter is not installed

on the drill attachment or the bur adapter

is damaged and not making contact with the

cutting accessory.

Make sure the electrical bur adapter is in-

stalled properly. See the

 To Install the Tempo-

ral Bone Insert and Connect Surgical Instru-

ments

 section. Replace the adapter if damage 

is apparent. See the 

Accessory Information

section.

The ground cable with hook-and-loop tape 

is not installed on the drill body or is not 

making contact with the drill body.

Make sure the ground cable with hook-and-

loop tape is installed properly.

See the 

To Install the Temporal Bone Insert

and Connect Surgical Instruments

 section.

The temporal bone insert is out of cali-

bration.

Recalibrate the bone insert. See the 

To

Perform T-Bone Calibration

 section.

Summary of Contents for S-00024

Page 1: ...USER MANUAL PHACON TEMPORAL BONE TRAINER BILATERAL...

Page 2: ...DISCOVER THE INTRODUCTION INTO THE PHACON TRAINER your laptop with navigation software your tracking camera your holder tray with head your trackers your patient your pointer...

Page 3: ...FORMATION 8 TRAINER COMPONENTS 9 INSTRUCTIONS 11 LAPTOP DISPLAY SCREEN 12 BUTTONS 13 OPERATE THE TRAINER 15 Enter the application 18 Perform setup for navigation 19 Perform patient calibration 20 Perf...

Page 4: ...rument tracker set for ENT surgery RE 00005 DESCRIPTION REF 3 instrument linkage cable RE 00009 1 notebook with PHACON simulation software RE 00011 1 USB 3 0 hub ZS 00006 1 computer mouse RE 00012 1 p...

Page 5: ...d lever lock to fasten the head to the ball joint mounting plate 3 Rotate the head to the desired position Rotate the ball joint lever lock to secure the head in place 4 Connect the camera USB extensi...

Page 6: ...pecific panel type C Camera Line of Sight Used to display the image within the navigation camera s line of sight D Simulation Time Used to display the start time of the simulation the current time and...

Page 7: ...e software German English Reset the default settings for brightness contrast and slice position in the 4 windowpane in the simulation mode Close and exit the program End the SIMULATION or training ses...

Page 8: ...click on the PHACON Trainer icon located on the laptop computer desktop to log into the training program ENTER THE APPLICATION 2 Type in your user name and password If you do not have an account you...

Page 9: ...fields are optional 4 From the MAIN MENU screen click the SIMULATION button Fig 3 Fig 3 Click on the SETUP button to perform instrument and patient calibration for optimal navigation results See To P...

Page 10: ...of the pointer drill or manual instrument The five calibration points are visible on the Temporal Bone Patient and within the 3 D windowpane during calibration Sequence numbers and a progress bar are...

Page 11: ...atient Make sure the tracking shapes are facing the camera and are within the camera line of sight 3 After visual and audible feedback indicates successful data acq uisition change the angle of the to...

Page 12: ...middle mouse button may also be used When using navigation the CT image slice is selected automatically based on the tip location of the pointer drill or manual instrument To view different slices of...

Page 13: ...he 3 D windowpane Fig 11 ADJUST THE OPACITY VALUE OF AN ANATOMICAL STRUCTURE The opacity of a specific anatomical structure may be changed to provide better visualization within and behind the Patient...

Page 14: ...ING which indicates that the recording has been started 2 To end the recording clickon the button RECORDING 3 To save the recorded video click on the CLASSIC but ton to change to the simulation mode N...

Page 15: ...ND button to end the training session and return to the MAIN MENU screen END THE SIMULATION OR TRAINING SESSION VIEW THE TRAINING SESSION RESULTS 1 From the MAIN MENU screen click on the RESU LTS butt...

Page 16: ...r group results Fig 20 1 Click on the ADD GRAPH button to compare your results with other group results Fig 20 2 From the COMPARISON WITH OTHER GROUPS dialog box clickon the GROUP drop down list and s...

Page 17: ...thin the head using a dry brush and an external vacuum source Make sure the spillways are clear to allow water to low freely during the use of irrigation 5 Wipe all the surfaces of the head including...

Page 18: ...s in the left column of the NVIDIA Control Panel 3 In the Global Settings klick on High performance NVIDIA processor 4 If the Apply button does not appear please select High performance NVIDIA process...

Page 19: ...e head in the free space on the top of the holder tray Now you can close the case lid top gently TRAINER STORAGE HOW TO PACK THE TRANSPORT CASE MESSAGE CAUSE ACTION The LOGIN password or user name is...

Page 20: ...ed with accessories not manufactured by PHACON or if repairs are performed by any party other than authorized PHACON repair personnel Except as set forth above phacon makes no warranties with respect...

Page 21: ...Germany tel 49 0 3 41 47 83 97 32 fax 49 0 3 41 47 83 91 74 www phacon de email info phacon us 4470 Chamblee Dunwoody Rd Suite 463 Dunwoody GA 30338 USA tel 1 833 2 PHACON 2 742266 fax 470 243 0494 w...

Reviews: