PERFORM PATIENT CALIBRATION
NOTES:
• Calibration is performed to align the skull model coordinate system with the coordinate system of the
pointer, drill, or manual instrument.
• The seven calibration points are visible on the Sinus Patient and within the 3-D windowpane during
calibration.
• Sequence numbers and a progress bar are provided in the Patient Calibration dialog window as visual
feedback during the calibration procedure.
1. Click on the RECALIBRATE button to begin the calibration process (
Fig. 6
).
2. Hold the tip of the pointer on the first calibration point of the Patient as indicated in the 3-D windowpane.
3. After visual and audible feedback indicates successful data acquisition, relocate the tip of the pointer
to the next calibration point as indicated in the 3-D windowpane.
4. Change the location of the pointer seven times. An INFORMATION MESSAGE pop-up will appear to
indicate the successful completion of the calibration procedure.
5. Click on the START/RESUME button to begin or continue the simulation, respectively, if desired.
Fig.: 6