8
PRECAUTIONS FOR SAFE OPERATION
Never install the unit in a location where corrosive materials such as sulphur compounds are
likely to be generated (e.g. near a drainage facility).
Corrosion of the copper pipes may result in
the deterioration and consequently the failure of the cooling unit.
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit.
A
multiple-tap may cause fire resulting from abnormal heating.
Do not climb on top of the unit or put any objects on the unit.
Falling from the unit may cause
injury; falling objects may cause damage to the unit.
Never store corrosive substances such as acids or alkalis in this unit except in a sealed
container.
These may be harmful to your health and may cause corrosion of internal components or
electrical parts.
Check the settings when restarting operation after a power failure or after turning the power off.
The settings may have changed as a result of stopping the unit. Stored items inside the unit may be
adversely affected when operation is resumed if the settings have changed.
To ensure the safety of the service engineer, submit a safety check sheet with the required
items filled out.
This is provided as the photocopiable “Safety Check Sheet” at the end of these
operating instructions
Use designated parts for parts replacement.
Using an incorrect part may cause fire.
Do not give strong shock or vibration during movement or use.
The piping may be damaged,
causing a fire.
Flammable and explosive product.
The unit contains flammable refrigerant. Consult repair
manual/owner’s guide before attempting to install or service this product. All safety precautions must
be followed.
Flammable and explosive product.
Dispose of properly in accordance with federal or local
regulations. Flammable refrigerant used.
To prevent frostbite, wear
protective gloves when handling
frozen items in the chamber.
Too
much frost may cause
chamber temperature rise
resulting from incomplete door
close.
Clean the filter about every once a month.
A dusty filter may cause poor cooling performance.
Please remove the frost on the air intake.
When removing, please use the accessory's stick for air
intake port cleaning.
CAUTION